1. Introducció i finalview

The Gilson Pipetman F100 is a fixed-volume, air-displacement pipette designed for precise liquid handling in laboratory settings. This model is specifically calibrated for a 100µl volume range, ensuring consistent and reliable results for routine testing, clinical diagnostics, and quality control applications. Its fixed-volume mechanism eliminates the risk of volume-selection errors during standard operation.

While primarily fixed, the pipette's volume can be adjusted by qualified users in a laboratory environment to accommodate for specific fluid properties, such as dense or viscous liquids. Once reset, the new volume remains fixed and cannot be accidentally altered during use.

Gilson Pipetman F100 Fixed-Volume Pipette, showing the main body, plunger, and tip ejector, with a disposable tip next to it.

Figura 1: The Gilson Pipetman F100 fixed-volume pipette. This image displays the blue main body with the 'Pipetman' logo, a yellow plunger mechanism visible at the top, and a white tip ejector rod extending downwards. A clear disposable pipette tip is shown separately, ready for attachment.

2. Configuració

  1. Desembalatge: Carefully remove the Pipetman F100 from its packaging. Inspect the pipette for any visible damage that may have occurred during transit.
  2. Inspecció inicial: Ensure all components are present and securely attached. The pipette should feel solid and free of loose parts.
  3. Tip Attachment: Select a high-quality, compatible pipette tip. Firmly press the pipette shaft into the opening of the tip with a slight twisting motion until a secure seal is formed. Avoid excessive force, which can damage the tip cone or affect accuracy.
  4. Volume Verification (Optional): Although fixed, it is recommended to perform an initial gravimetric check with distilled water to confirm the pipette's accuracy, especially if it is new or has been stored for an extended period. Refer to standard laboratory procedures for gravimetric calibration.

3. Instruccions de funcionament

The Gilson Pipetman F100 is designed for straightforward operation due to its fixed-volume nature. Follow these steps for accurate liquid transfer:

  1. Attach a Tip: Securely attach a new, clean pipette tip to the tip cone as described in the Setup section.
  2. Pre-wetting (Recommended): For optimal accuracy, especially with volatile liquids, aspirate and dispense the liquid once or twice before transferring the actual sample. This saturates the air space in the tip with vapor, reducing evaporative loss during the actual transfer.
  3. Aspiració:
    • Depress the plunger smoothly to the first stop.
    • Immerse the tip into the liquid to a depth of 2-3 mm.
    • Slowly release the plunger to allow the liquid to be drawn into the tip. Wait 1-2 seconds after the plunger reaches its resting position to ensure the full volume is aspirated.
    • Withdraw the tip from the liquid, touching it against the edge of the reservoir to remove any excess droplets.
  4. Dispensació:
    • Place the tip against the inner wall of the receiving vessel at a slight angle.
    • Smoothly depress the plunger to the first stop, expelling the majority of the liquid.
    • Continue depressing the plunger to the second stop (blow-out) to ensure complete delivery of the remaining liquid.
    • While holding the plunger at the second stop, withdraw the pipette from the vessel.
  5. Tip Ejection: Press the tip ejector button to discard the used tip into an appropriate waste container. Never reuse tips.

4. Manteniment

Proper maintenance ensures the longevity and accuracy of your Gilson Pipetman F100.

  • Neteja diària: Netegeu l'exterior de la pipeta amb un drap suauampened with 70% ethanol or a mild detergent solution. Avoid harsh solvents.
  • Descontaminació: If the pipette comes into contact with hazardous materials, follow established laboratory decontamination protocols. The lower part of the pipette (tip ejector and tip cone) can often be autoclaved or disassembled for thorough cleaning, consult the full manufacturer's manual for specific instructions.
  • Emmagatzematge: Store the pipette vertically in a pipette stand to prevent contamination and damage to the tip cone. Ensure it is stored in a clean, dry environment away from direct sunlight and extreme temperatures.
  • Comprovació periòdica de calibratge: Even though it's a fixed-volume pipette, periodic gravimetric calibration checks (e.g., every 3-6 months or as per laboratory SOPs) are crucial to confirm continued accuracy and precision. If the pipette consistently falls outside acceptable limits, it may require servicing or recalibration by a qualified technician.

5. Solució De Problemes

If you encounter issues with your Pipetman F100, consider the following common problems and solutions:

ProblemaCausa possibleSolució
Inaccurate or inconsistent dispensing
  • Loose or incompatible tip
  • Damaged tip cone
  • Contamination inside the pipette
  • Improper pipetting technique (e.g., too fast aspiration/dispensing)
  • Temperature differences between pipette, liquid, and environment
  • Ensure tip is securely attached and compatible.
  • Inspect tip cone for cracks or damage; replace if necessary.
  • Perform thorough cleaning and decontamination.
  • Review operating instructions for correct technique.
  • Allow pipette and liquids to equilibrate to room temperature.
  • Perform a gravimetric calibration check.
Líquid que goteja de la punta
  • Loose or incompatible tip
  • Damaged tip cone or seal
  • Contamination affecting seal
  • Re-attach tip firmly or try a new tip.
  • Inspect tip cone for damage.
  • Clean the tip cone and shaft.
  • If problem persists, contact Gilson support.
Plunger movement is stiff or sticky
  • Falta de lubricació
  • Contamination in the plunger mechanism
  • Consult the full manufacturer's manual for lubrication instructions.
  • Disassemble and clean the plunger assembly if comfortable, or send for professional servicing.

For issues not listed here or if problems persist, refer to the comprehensive Gilson Pipetman F100 user manual or contact Gilson technical support.

6. Especificacions

CaracterísticaDetall
ModelPipetman F100
Interval de volum100 µL (Fixed)
Tipus de pipetaAir-displacement, Fixed-Volume
FabricantGilson
Dimensions del paquet12.5 x 4.3 x 1.5 polzades (31.75 x 10.92 x 3.81 cm)
Pes del paquet8 unces (0.5 lliures)
ASINB008TPWFLA

7. Garantia i Suport

For detailed warranty information, service, or technical support regarding your Gilson Pipetman F100, please refer to the official Gilson website or contact Gilson customer service directly. Ensure you have your product model number (F100) and any purchase details available when seeking support.

Visit Gilson's Official Weblloc

© 2023 Gilson. All rights reserved. This manual is for informational purposes only.