Gain Express 8403_FBACO

Gain Express Dissolved Oxygen Meter

Model: 8403_FBACO

1. Introducció

1.1 Producte acabatview

The Gain Express Dissolved Oxygen Meter is a high-precision instrument designed for measuring dissolved oxygen (DO) levels, temperature, and barometric pressure. It is equipped with advanced features for accurate and reliable readings in various environments.

1.2 Característiques clau

  • High-precision DO measurement (%, ppm/mg/L)
  • Large LCD display for simultaneous readings of DO, temperature, and barometric pressure
  • Compensació automàtica de temperatura (ATC)
  • Manual salinity adjustment
  • 99-data memory with real-time clock
  • Funció de retenció de dades
  • Indicador de bateria baixa
  • Selectable auto power-off (20-120 minutes)
  • Durable Clark-type electrode with 11.48 ft cable and extended lifespan with proper maintenance

1.3 Aplicacions

This dissolved oxygen meter is suitable for a wide range of applications, including:

  • Aquaculture (fish farming, aquariums)
  • Industrial water quality monitoring
  • Environmental research and monitoring
  • Finalitats educatives
Dissolved Oxygen Meter showing various applications like aquaculture, hydroponics, and laboratory testing.

Image 1.1: The Dissolved Oxygen Meter is used in diverse settings such as aquaculture, hydroponics, and scientific laboratories.

Vídeo 1.1: Un overview of the Dissolved Oxygen Meter, its features, and various applications in water quality monitoring.

2. Contingut del paquet

En obrir el paquet, comproveu que tots els elements enumerats a continuació hi siguin presents i estiguin en bon estat:

  • Dissolved Oxygen Meter Unit
  • DO Probe with 11.48 ft (3.5m) Cable
  • 4 x piles AAA
  • 5 x Teflon Membranes
  • 5 x juntes tòriques
  • 2 x Electrolyte Bottles
  • Manual d'instruccions en anglès
  • Sturdy Carrying Case
Contents of the Dissolved Oxygen Meter kit, including the meter, probe, batteries, membranes, electrolyte, manual, and carrying case.

Image 2.1: All components included in the Dissolved Oxygen Meter kit.

3. Components mésview

3.1 Meter Unit

The main meter unit features a large LCD display and several function keys for operation.

  • Pantalla LCD: Shows dissolved oxygen readings, temperature, barometric pressure, time, and various indicators.
  • Tecles de funció:
    • SET: Entra en el mode de configuració o confirma els paràmetres.
    • Mode: Switches between measurement modes or menu options.
    • CAL/Esc: Initiates calibration or exits current menu/setting.
    • MEM: Accesses memory records.
    • HLD/Rec: Activates data hold or recording function.
    • MR/AV: Recalls memory data or calculates average.
  • Probe Port: Connector for the DO probe.
Diagram of the Dissolved Oxygen Meter showing the LCD display, function keys, and probe port.

Image 3.1: Labeled diagram of the Dissolved Oxygen Meter unit.

3.2 DO Probe

The Clark-type DO probe is responsible for sensing dissolved oxygen levels. It connects to the main unit via a cable.

  • Consell del sensor: Contains the sensitive membrane and electrodes for DO measurement.
  • Cable: Connects the probe to the meter unit.
Close-up image of the Dissolved Oxygen probe with its cable.

Image 3.2: The Dissolved Oxygen probe.

4. Configuració

4.1 Instal·lació de la bateria

  1. Locate the battery compartment on the back of the meter unit.
  2. Obriu la coberta del compartiment.
  3. Introduïu 4 piles AAA, assegurant-vos de la polaritat correcta (+/-).
  4. Tanqueu bé la tapa del compartiment de la bateria.

4.2 Connexió de la sonda

  1. Align the connector of the DO probe cable with the probe port on the top of the meter unit.
  2. Gently push and twist the connector clockwise until it is securely fastened.

4.3 Electrolyte Preparation and Membrane Installation

The DO probe requires an electrolyte solution and a membrane cap for proper function. Refer to the detailed instructions in the included manual for precise steps on preparing the electrolyte solution and installing the Teflon membrane and O-rings onto the probe cap.

5. Instruccions de funcionament

5.1 Encès/Apagat

  • Premeu el botó SET button to power on the meter.
  • Per apagar-lo, manteniu premut el botó SET button until the display turns off, or allow the auto power-off function to activate.

5.2 Modes de mesura

The meter can display dissolved oxygen in percentage (%), parts per million (PPM), or milligrams per liter (mg/L).

  • Premeu el botó MODE button to cycle through the available DO measurement units.

5.3 Mesura de la temperatura

The meter automatically measures and displays temperature alongside DO readings. The temperature unit can be switched between Celsius and Fahrenheit.

  • Refer to the full manual for instructions on changing the temperature unit.

5.4 Salinity Adjustment

For accurate DO readings in saline water, manual salinity compensation is required.

  • Enter the setup menu using the SET button and navigate to the salinity setting.
  • Adjust the salinity value according to your sample's known salinity.

5.5 Barometric Pressure Input

Barometric pressure affects dissolved oxygen solubility. The meter allows for manual input of barometric pressure for compensation.

  • Access the barometric pressure setting in the setup menu to input the local barometric pressure.

5.6 Funció de retenció de dades

  • Premeu el botó HLD/Rec button briefly to freeze the current reading on the display.
  • Press it again to release the hold.

5.7 Funció de memòria

The meter can store up to 99 data sets, each with a real-time clock stamp.

  • To save a reading, press and hold the HLD/Rec button until 'REC' appears.
  • To recall saved data, press the MEM button. Use the navigation buttons (if available, or refer to manual) to scroll through records.
Dissolved Oxygen Meter displaying operating features like manual salinity, memory, IRDA, and real-time clock.

Image 5.1: Operating features of the Dissolved Oxygen Meter.

Dissolved Oxygen Meter with an illustration of 99 memory slots.

Image 5.2: The meter features 99 data memory with real-time clock.

Dissolved Oxygen Meter display showing a low battery indicator.

Image 5.3: Low battery indicator on the meter's display.

6. Calibració

Regular calibration is essential for maintaining the accuracy of your Dissolved Oxygen Meter. The meter supports both air saturation and zero oxygen calibration methods.

6.1 Air Saturation Calibration

This method calibrates the probe in air-saturated water or in humid air, representing 100% saturation.

  1. Ensure the probe is clean and has a properly installed membrane and fresh electrolyte.
  2. Place the probe in a humid air environment (e.g., in a bottle with a small amount of water, ensuring the probe tip is not submerged) or in air-saturated water.
  3. Premeu el botó CAL botó per entrar al mode de calibratge.
  4. Follow the on-screen prompts to confirm the air saturation point.

6.2 Calibració d'oxigen zero

This method calibrates the probe to a zero oxygen level, typically using a zero-oxygen solution.

  1. Prepare a zero-oxygen solution (e.g., using sodium sulfite).
  2. Immerse the probe into the zero-oxygen solution.
  3. Premeu el botó CAL button and follow the on-screen prompts to confirm the zero oxygen point.

For detailed step-by-step calibration procedures, including specific values and stability requirements, please refer to the comprehensive instruction manual provided with your device.

7. Manteniment

Proper maintenance ensures the longevity and accuracy of your Dissolved Oxygen Meter.

7.1 Cura de la sonda

  • Always keep the probe tip moist when not in use. Use the protective cap filled with a small amount of storage solution or clean water.
  • Clean the probe regularly with distilled water to remove any residue. Avoid abrasive materials.

7.2 Membrane Replacement

The Teflon membrane is a consumable part and needs periodic replacement, especially if damaged or if readings become unstable.

  • Carefully unscrew the membrane cap from the probe.
  • Remove the old membrane and O-ring.
  • Install a new Teflon membrane and O-ring, ensuring it is smooth and free of wrinkles or air bubbles.
  • Refill the cap with fresh electrolyte solution before screwing it back onto the probe.

7.3 Electrolyte Solution Replacement

The electrolyte solution inside the probe cap should be replaced periodically (e.g., every 1-3 months or when readings become sluggish).

  • Follow the membrane replacement procedure to access the electrolyte.
  • Discard the old solution and refill with fresh electrolyte from the provided bottles.

7.4 Emmagatzematge

  • Store the meter and probe in the provided carrying case when not in use.
  • Remove batteries if the meter will not be used for an extended period to prevent leakage.
  • Emmagatzemar en un lloc fresc i sec, lluny de la llum solar directa i de temperatures extremes.

8. Solució De Problemes

If you encounter issues with your Dissolved Oxygen Meter, refer to the table below for common problems and solutions.

ProblemaCausa possibleSolució
El comptador no s'encénBateries gastades o mal instal·ladesCanvieu les piles, assegurant-vos de la polaritat correcta.
Lectures inexactes o inestables
  • Probe not calibrated
  • Damaged or dirty membrane
  • Old or contaminated electrolyte
  • Air bubbles on membrane
  • Incorrect salinity/pressure compensation
  • Perform calibration (air saturation/zero oxygen).
  • Inspect and clean/replace membrane.
  • Replace electrolyte solution.
  • Gently tap probe to dislodge bubbles.
  • Verify and adjust salinity/pressure settings.
"Low Battery" indicatorLes bateries estan baixesSubstituïu les 4 piles AAA.
Cap resposta dels botonsMeter frozen or malfunctionRemove and reinsert batteries to reset the meter. If problem persists, contact support.

9. Especificacions

ParàmetreIntervalPrecisióResolució
Oxigen dissolt (OD)0~199.9% / 0~19.99 PPM / 0~19.99 mg/L±1.5% FS0.1% / 0.01 PPM / 0.01 mg/L
Temperatura0 ~ 50 ° C (32 ~ 122 ° F)±0.3 °C0.1 °C
Compensació de salinitat0.0~50.0 pptN/A0.1 ppt
Pressió baromètrica500~1499 mmHg / 66.6~199.9 KPaN/A1 mmHg / 0.1 KPa

Especificacions generals

  • Número de model: 8403_FBACO
  • Font d'alimentació: 4 piles AAA (incloses)
  • Memòria: 99 conjunts de dades
  • Entorn de funcionament: 0-50 °C (32-122 °F)
  • Probe Cable Length: 11.48 peus (3.5 m)
  • Pes de l'article: 1 lliures
  • Dimensions del paquet: 6.97 x 5.75 x 2.72 polzades

10. Garantia i Suport

10.1 Informació de la garantia

This product is covered by a standard manufacturer's warranty against defects in materials and workmanship. Please retain your proof of purchase for warranty claims. Specific warranty duration and terms may vary; refer to the warranty card included in your package or contact customer support for details.

10.2 Atenció al client

For technical assistance, troubleshooting beyond this manual, or warranty inquiries, please contact Gain Express customer support. Contact information can typically be found on the product packaging, the official Gain Express weblloc web o a través del vostre distribuïdor.