Elitech GSP-6-PTE-US
Elitech GSP-6-PTE Ultra Low Digital Temperature Data Logger User Manual
Model: GSP-6-PTE-US
Marca: Elitech
1. Introducció
The Elitech GSP-6-PTE Ultra Low Digital Temperature Data Logger is a high-precision device designed for monitoring temperature and humidity, particularly suitable for medical refrigerators, vaccine storage, and cold chain management. It meets current CDC / VFC requirements and additional recommendations for vaccine storage and monitoring, and is FDA 21 CFR Part 11 Compliant. This manual provides comprehensive instructions for setting up, operating, maintaining, and troubleshooting your data logger.

Figure 1: Elitech GSP-6-PTE Data Logger
2. Producte acabatview
The GSP-6-PTE data logger features a large LCD screen for clear value visibility, a high-precision external probe, and dual power options (battery and USB). Its robust design ensures reliable performance in demanding environments.
Característiques principals:
- Vaccine Storage Compliance: Designed to meet CDC / VFC requirements and FDA 21 CFR Part 11.
- Fàcil instal·lació: Built-in magnet for direct attachment to refrigerators. Separable probe eliminates the need for pre-cooling after device movement.
- Sonda d'alta precisió: Accurately senses small temperature changes with a wide measurement range (-121°F to 302°F) and humidity (10%~99%RH).
- Doble potència: Operates on battery or USB power.
- Value Visibility: Current temperature, maximum, and minimum values are easily viewable on the large LCD screen.
- Large Capacity Data Storage: Stores up to 16,000 data points, allowing continuous recording for extended periods.

Figure 2: GSP-6-PTE Component Diagram
3. Configuració
3.1 Desembalatge i inspecció inicial
Upon receiving your Elitech GSP-6-PTE, carefully unpack all components and inspect for any damage. The package should include:
- Elitech GSP-6-PTE Data Logger
- Sonda de sensor extern
- Certificat de calibratge
- Manual d'usuari (aquest document)
3.2 Alimentació del dispositiu
The device can be powered by its internal lithium metal battery (included) or via a USB connection.
- Potència de la bateria: The device comes with a pre-installed battery. Ensure it is properly seated.
- Potència USB: Connect the device to a USB port using a standard USB cable. This can also be used to power the device when the battery is low or for continuous operation.

Figure 3: Dual Power (Battery & USB)
3.3 Connecting the External Probe
The GSP-6-PTE utilizes an external probe for temperature and humidity measurement. Ensure the probe is securely connected to the device's designated port. If the screen displays "Err", it indicates the probe is not properly connected.
Video 1: Demonstrates connecting the probe and navigating the display. If "Err" appears, ensure the probe is properly connected.

Figure 4: External Probe Placement
3.4 Software Installation (Elitechlog)
To fully utilize the data logging capabilities, including data export and advanced settings, install the Elitechlog software on your computer (Windows or Mac OS). The software is typically provided on a mini-CD or can be downloaded from the Elitech official weblloc.
- Connect the data logger to your computer via USB.
- Follow the on-screen instructions to install the Elitechlog software.
- Once installed, you can configure logging intervals, alarm limits, and export data.

Figure 5: Exporting Data via Elitechlog Software
4. Instruccions de funcionament
4.1 Navegació de la pantalla
The large LCD screen provides real-time data and various logged values. Press the main button on the device to cycle through different display modes:
- Interfície principal: Displays current temperature and humidity, along with the number of logging points.
- Valor màxim: Shows the maximum recorded temperature and humidity, along with the time it occurred.
- Valor mínim: Shows the minimum recorded temperature and humidity, along with the date it occurred.
- Valor mitjà: Displays the average temperature and humidity over the logging period.
- Upper Alarm Limit: Shows the configured upper temperature and humidity alarm thresholds.
- Lower Alarm Limit: Shows the configured lower temperature and humidity alarm thresholds.

Figure 6: Large Screen Display
4.2 Starting and Stopping Logging
Logging can be started and stopped via the Elitechlog software. Refer to the software's user guide for detailed instructions on setting logging intervals (e.g., 10 seconds to 24 hours) and starting/stopping data collection.
4.3 Data Export and Analysis
Once data has been collected, connect the data logger to your computer and use the Elitechlog software to export the data. The software allows you to:
- Query and store data.
- Generate chart comparisons for multiple devices.
- Switch between Celsius and Fahrenheit temperature units.
- Set audible and indicator alarm ranges.
- Export data graphs and tables in formats like PDF and Excel.
- Utilize email backup features.

Figure 7: Elitechlog Software Capabilities
5. Manteniment
5.1 Substitució de la bateria
The GSP-6-PTE uses a Lithium Metal battery. When the battery indicator on the display shows low power, replace the battery to ensure continuous operation and data integrity. Refer to the device's specific battery compartment for replacement instructions.
5.2 Neteja i emmagatzematge
Clean the device and probe with a soft, dry cloth. Avoid using abrasive cleaners or solvents. Store the data logger in a cool, dry place away from direct sunlight and extreme temperatures when not in use.
5.3 Cura de la sonda
Handle the external probe with care. Avoid bending or kinking the cable excessively. Ensure the probe tip remains clean for accurate readings. The probe is IP64 waterproof grade, but avoid prolonged submersion.
6. Solució De Problemes
6.1 "Err" Displayed on Screen
If the device displays "Err" for temperature or humidity, it indicates that the external probe is not properly connected or is faulty. Ensure the probe's jack is fully inserted into the device's port. If the issue persists after re-connecting, the probe may need replacement.
6.2 El dispositiu no respon
If the device is unresponsive, check the battery level. Connect the device to a USB power source to see if it powers on. If it still does not respond, try replacing the battery.
6.3 Data Not Exporting
Ensure the Elitechlog software is correctly installed and the data logger is properly connected to the computer via USB. Verify that the USB drivers are installed. Try restarting both the device and the software. Consult the Elitechlog software's help section for specific connectivity issues.
6.4 Lectures inexactes
If you suspect inaccurate readings, ensure the probe is clean and positioned correctly in the environment being monitored. Compare readings with a known calibrated thermometer. The Elitechlog software may offer calibration adjustment options if minor discrepancies are noted.
7. Especificacions
| Característica | Detall |
|---|---|
| Número de model | GSP-6-PTE-US |
| Marca | Elitech |
| Interval de mesura de temperatura | -121 °F a 302 °F (-85 °C a 150 °C) |
| Interval de mesura de la humitat | 10% ~ 99% HR |
| Precisió de la temperatura | ±0.5 °C |
| Capacitat d’emmagatzematge de dades | 16,000 points (Max) |
| Interval de registre | 10 seconds to 24 hours (configurable) |
| Font d'alimentació | 1 Lithium Metal battery (included), USB Power |
| Tipus de muntatge | Surface Mount (with built-in magnet) |
| Dimensions del paquet | 6.65 x 4.53 x 3.23 polzades |
| Pes de l'article | 11.68 unces |
| Primera data disponible | 15 d'octubre de 2021 |
8. Garantia i Suport
Elitech provides comprehensive customer support for its products. For warranty information, technical assistance, or further inquiries, please visit the official Elitech weblloc o poseu-vos en contacte amb el seu departament d'atenció al client.
Podeu trobar més informació i recursos de suport a Elitech Store on Amazon.
Ask a question about this manual
Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.