INSTRUKART AI-DP1-LCD

INSTRUKART AI-DP1-LCD Digital Differential Pressure Indicator User Manual

Model: AI-DP1-LCD | Brand: INSTRUKART

1. Introducció

The INSTRUKART AI-DP1-LCD is a digital differential pressure indicator designed for precise measurement in critical environments such as laminar airflow cabinets, clean rooms, biosafety cabinets, and Air Handling Units (AHU). This device serves as a modern replacement for traditional analog Magnehelic gauges, offering enhanced readability and functionality. It features a clear Graphic Backlit 128 x 64 LCD, an integrated high-speed differential pressure sensor, and a built-in buzzer alarm for process violation alerts with an acknowledgment key.

This manual provides essential information for the safe and efficient operation, installation, and maintenance of your AI-DP1-LCD Differential Pressure Indicator. Please read this manual thoroughly before using the device.

2. Informació de seguretat

Always adhere to local safety regulations and guidelines when installing and operating the device. Ensure that the power supply matches the specified requirements (24V DC). Do not expose the device to corrosive gases or liquids. Only qualified personnel should perform installation and maintenance.

3. Contingut del paquet

En desembalar, comproveu que tots els elements que s'enumeren a continuació hi són i no estan danyats:

  • 1 Unit of INSTRUKART AI-DP1-LCD Differential Pressure Indicator
  • Adaptador d'alimentació
  • Manual d'instruccions (aquest document)
  • Factory Calibration Certificate with NABL Standards

4. Producte acabatview

The AI-DP1-LCD is designed for accurate differential pressure monitoring. Its robust construction and advanced features make it suitable for demanding industrial and scientific applications.

4.1 Característiques clau

  • Digital readout with Graphic Backlit 128 x 64 LCD.
  • Integrated high-speed differential pressure sensor.
  • Built-in high and low buzzer alarm for process set parameter violation with acknowledgment key.
  • Unit conversion capability (mm.w.c. & Pascals) - Note: Unit selection is factory fixed for this model (-180 to 180 Pascals).
  • Offset provision to meet audit requirements.
  • 30 mm depth for cleanroom modular wall fitment.
  • Isolated RS485 Modbus communication for BMS / SCADA / PLC integration.

4.2 Producte Views

INSTRUKART AI-DP1-LCD Differential Pressure Indicator, angled view

Figura 1: Angle view of the INSTRUKART AI-DP1-LCD Differential Pressure Indicator.

Davant view of the INSTRUKART AI-DP1-LCD Differential Pressure Indicator

Figura 2: Frontal view of the AI-DP1-LCD, showing the display and 'ACK' button.

Lateral i frontal view comparison of the INSTRUKART AI-DP1-LCD

Figure 3: Comparison of the side and front views, highlighting the slim profile and rear connections.

5. Configuració i instal·lació

The AI-DP1-LCD is designed for modular wall fitment in cleanroom environments.

5.1 Muntatge

The instrument enclosure has dimensions of 100mm (W) × 90mm (H) × 30mm (D), allowing for flush mounting into cleanroom modular walls. Ensure adequate space for rear connections.

5.2 Connexions de pressió

Connect the pressure tubing to the two 4mm stainless steel hose nipples located at the back of the unit. Ensure connections are secure to prevent leaks. The device is designed for use with air or non-corrosive gases.

5.3 Connexió d'alimentació

Connect the provided power adaptor to the device's power input. The typical power requirement is 24V DC.

5.4 Communication (Optional)

For integration with Building Management Systems (BMS), SCADA, or PLC, utilize the isolated RS485 communication port with Modbus protocols. Refer to the Modbus communication protocol documentation for detailed setup instructions (not included in this manual).

6. Instruccions de funcionament

6.1 Encès

Once powered on, the Graphic Backlit LCD will display the current differential pressure reading. The display also shows the date and time.

6.2 Reading Display

The LCD shows the differential pressure value in Pascals. This model has a factory-fixed unit selection for Pascals.

6.3 Funció d'alarma

The device features a built-in high and low buzzer alarm. If the measured differential pressure exceeds or falls below the pre-set parameters, the buzzer will activate to indicate a process violation.

6.4 Alarm Acknowledgment

To acknowledge and silence an active alarm, press the 'ACK' button located on the front panel below the display. Acknowledging the alarm does not resolve the underlying issue; it only silences the buzzer.

7. Manteniment

7.1 Neteja

Wipe the front stainless steel plate and LCD screen with a soft, damp cloth. Do not use abrasive cleaners or solvents, as these may damage the surface or display. Ensure no liquid enters the device.

7.2 Calibració

The device comes with a Factory Calibration Certificate with NABL Standards. Regular re-calibration may be required depending on application and regulatory requirements. Contact INSTRUKART for calibration services.

8. Solució De Problemes

ProblemaCausa possibleSolució
No display/Device not powering onNo hi ha font d'alimentació o volum incorrectetage.Check power adaptor connection and ensure 24V DC supply.
Lectura de pressió incorrectaLeaky pressure connections or blocked hose nipples.Verify all hose connections are secure. Check for any blockages in the tubing or nipples.
Alarm constantly activeDifferential pressure outside set parameters.Investigate the environmental conditions causing the pressure deviation. Acknowledge the alarm by pressing 'ACK'.
Communication issues (RS485)Incorrect wiring or Modbus configuration.Verify RS485 wiring and Modbus parameters. Consult Modbus documentation or INSTRUKART support.

9. Especificacions tècniques

ParàmetreDetall
Interval de mesura-180 to 180 Pascals
Precisió±0.5% de l'escala completa
MostraGraphic Backlit 128 x 64 LCD
Maximum Over Pressure4 Times the rated pressure
SensorIntegrated High Speed Differential Pressure Sensor
Font d'alimentació24V DC Typical
AlarmesBuilt-in High & Low Buzzer Alarm with Acknowledgment Key
ConstruccióM.S. Powder Coated Body (Back) with Stainless Steel Front Flush Plate
ComunicacióIsolated RS485 communication, Modbus Protocols
Connexió de pressió2 x 4mm Stainless Steel Hose Nipples at the back
MitjanaAir or Non-Corrosive Gases
Dimensions (A×A×P)100 mm × 90 mm × 30 mm
Selecció d'unitatsFactory Fixed for Pascals
Pes de l'article1.76 lliures (aprox. 0.8 kg)

10. Aplicacions

The INSTRUKART AI-DP1-LCD is suitable for a wide range of critical environments requiring precise differential pressure monitoring:

  • Infection Rooms
  • Hospital Operation Theatre (OT) rooms
  • Isolation Rooms (especially where negative pressure is required)
  • Laminar Air Flow Cabinets
  • Pharmaceutical Clean Rooms
  • Sistemes HVAC
  • Bio Safety Cabinets
  • Air Handling Units (AHUs)
INSTRUKART AI-DP1-LCD in an isolation room setting

Figure 4: The AI-DP1-LCD installed in an isolation room, demonstrating its application in critical healthcare environments.

Collage of various applications for the INSTRUKART AI-DP1-LCD, including isolation rooms, pharmaceuticals, AHUs, and HVAC systems

Figure 5: Visual representation of diverse applications for the differential pressure indicator, including cleanrooms, HVAC, and pharmaceutical settings.

11. Garantia i Suport

For warranty information, technical support, or service inquiries, please contact INSTRUKART directly. Refer to the product packaging or INSTRUKART's official weblloc web per obtenir les dades de contacte més actualitzades.

© 2024 INSTRUKART. All rights reserved. Information subject to change without notice.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.