1. Introducció
The Micsig CP503B is a high-frequency AC/DC current probe designed for precise current measurements with oscilloscopes. It offers a bandwidth of DC~50MHz and current ranges of 5A/30A, making it suitable for various electrical testing applications. This manual provides essential information for the proper setup, operation, and maintenance of your CP503B current probe.

Figure 1: Micsig CP503B High Frequency AC/DC Current Probe.
2. Producte acabatview i Característiques clau
The Micsig CP503B current probe is engineered for accuracy and ease of use in demanding measurement environments.
Característiques principals:
- Ample de banda: DC~50MHz for comprehensive signal analysis.
- Interval actual: Selectable 5A/30A for diverse measurement needs.
- DC Accuracy (Typical): ±1% for reliable readings.
- Selecció de doble rang: Facilitates accurate measurement of both low and high currents.
- Auto Degaussing and Zero: Simplifies setup and ensures measurement precision.
- Indicador de sobrecàrrega: Provides visual alert for current exceeding the selected range.
- Standard BNC Interface: Ensures compatibility with all standard oscilloscopes.
- Compact and Exquisite Design: Ergonomic for comfortable handling.
Aplicacions típiques:
- Disseny de transport de vehicles elèctrics
- Disseny d'alimentació commutada
- Electronic engineering experiments
- Semiconductor device design
- Disseny avònic
- Disseny inversor/transformador
- Disseny de balast electrònic
- Industrial control and consumer electronics design
- Disseny accionat per motor
- Disseny experimental d'electrònica de potència i accionament elèctric

Figure 2: Micsig current probe connected to an oscilloscope.
3. Instruccions de configuració
Follow these steps to set up your Micsig CP503B current probe for operation.
- Connectar-se a l'oscil·loscopi: Connect the BNC connector of the current probe module to an input channel on your oscilloscope.
- Connexió d'alimentació: Connect the probe to a DC 5V 3A power source using the provided cable. This typically involves a USB power adapter or a compatible power supply.
- Encesa inicial: Once powered, the probe will perform an automatic degaussing and zero adjustment. Allow a few seconds for this process to complete.
- Configuració de l'oscil·loscopi: Configure your oscilloscope input channel to match the probe's output sensitivity (e.g., 1V/2A or 1V/10A depending on the selected range). Refer to your oscilloscope's manual for specific input scaling settings.

Figura 3: Detallada view of the current probe head.
4. Instruccions de funcionament
Proper operation ensures accurate and safe current measurements.
4.1. Range Selection
The CP503B offers two current ranges: 5A and 30A. Select the appropriate range based on the expected current magnitude of your measurement.
- Premeu el botó Interval button on the probe module to toggle between 5A and 30A.
- An indicator LED will illuminate next to the selected range (5A or 30A).

Figure 4: Range selection buttons and indicator LEDs.
4.2. Degaussing and Auto Zero Setting
To ensure measurement accuracy, especially for DC current measurements, perform a degaussing and zero adjustment before each measurement or if the probe has been exposed to strong magnetic fields.
- Ensure no current is flowing through the probe jaws.
- Manteniu premut el botó Zero button on the probe module until the indicator light flashes. This initiates the degaussing and auto-zero process.
- Release the button. The probe is now zeroed and ready for measurement.

Figure 5: Zero button for degaussing and auto-zero.
4.3. Making a Measurement
To measure current, open the probe jaws and clamp them around the conductor carrying the current you wish to measure. Ensure the jaws are fully closed around a single conductor. The current waveform will be displayed on the connected oscilloscope.
4.4. Overload Indication
If the measured current exceeds the selected range, the overload indicator LED on the probe module will flash. In such cases, switch to a higher current range (e.g., from 5A to 30A) or ensure the current does not exceed the probe's maximum input rating.
5. Application Performance Examples
The CP503B is capable of capturing various current signals with high fidelity.

Figure 6: High-frequency AC current signal (30A, 0.1V/A).

Figure 7: Surge current waveform during power adapter startup (6A, 0.5V/A).
6. Manteniment
To ensure the longevity and accuracy of your Micsig CP503B current probe, follow these maintenance guidelines:
- Neteja: Use a soft, dry cloth to clean the probe's exterior. Do not use abrasive cleaners or solvents. Ensure no liquids enter the probe housing or BNC connector.
- Emmagatzematge: Store the probe in a clean, dry environment, away from extreme temperatures, humidity, and strong magnetic fields. The original carrying case is recommended for protection during transport and storage.
- Cura del cable: Avoid sharp bends or kinks in the cable. Do not pull the cable to disconnect the probe; always grasp the connector.
- Jaw Integrity: Periodically inspect the probe jaws for any signs of damage or misalignment. Ensure they close completely and smoothly.
7. Solució De Problemes
This section addresses common issues you might encounter with your CP503B current probe.
- Sense senyal de sortida:
- Ensure the probe is properly powered (DC 5V 3A). Check power indicator lights.
- Verify the BNC connection to the oscilloscope is secure.
- Confirm the oscilloscope input channel is enabled and correctly configured (e.g., input impedance, vertical scale).
- Ensure the probe jaws are fully closed around the conductor.
- Inaccurate Readings / DC Offset:
- Perform a degaussing and auto-zero procedure (refer to Section 4.2).
- Check if the selected current range is appropriate for the measured current.
- Ensure the oscilloscope's vertical scale (V/div) is correctly set according to the probe's output sensitivity (e.g., 1V/2A or 1V/10A).
- Overload Indicator Flashing:
- The measured current exceeds the selected range. Switch to a higher current range (e.g., from 5A to 30A).
- If the indicator still flashes on the highest range, the current exceeds the probe's maximum rating. Do not attempt to measure currents beyond the probe's specifications to avoid damage.
- Probe Runs Warm:
- It is normal for active current probes to generate some heat during operation. If the probe becomes excessively hot or emits smoke/unusual odors, disconnect power immediately and cease use.
8. Especificacions
The following table details the technical specifications for the Micsig CP503B and CP1003B current probes.
| Paràmetre | CP503B | CP1003B |
|---|---|---|
| Model | CP503B | CP1003B |
| Ample de banda | 50 MHz | 100 MHz |
| Temps de pujada | ≤7ns | ≤3.5ns |
| Interval | 6A (2X)/30A (10X) | |
| Sensibilitat de sortida | 1V/2A (6A) 1V/10A (30A) | |
| DC Accuracy (typical) | ±1%±10mA (6A) ±1%±50mA (30A) | |
| Retard | <30ns (6A) <30ns (30A) | |
| Interval actual | 20mA~6Apk (6A) 50mA~30Apk (30A) | |
| Màx. Entrada actual | 30Apk, 60Apk-pk, 21.21Arms | |
| Soroll | ≤ 1.4 mArms (Bandwidth at 20 MHz, Range 30A, 10X) | |
| Màx. Voltage | CAT I 300V | |
| Màx. Vol. flotanttage | CAT I 300V | |
| Màx. Diàmetre del conductor | 5 mm | |
| Indicador de sobrecàrrega | Llum intermitent | |
| Font d'alimentació | DC 5V 3A | |

Figure 8: Technical specifications and characteristic curves.
9. Informació de la garantia
Micsig products are typically covered by a limited warranty against defects in materials and workmanship. The specific terms and duration of the warranty may vary by region and product. Please refer to the warranty card included with your product or visit the official Micsig weblloc per obtenir informació detallada sobre la garantia.
10. Informació de suport
For technical assistance, product inquiries, or service requests, please contact Micsig customer support through the following channels:
- Oficial Weblloc: Visit the Micsig official website for FAQs, software downloads, and contact forms.
- Suport per correu electrònic: Refer to your product packaging or the Micsig weblloc web per obtenir l'adreça de correu electrònic d'assistència adequada.
- Distribuïdors autoritzats: Contact your local Micsig distributor for regional support.