PINTECH N1070A

PINTECH N1070A High Voltage Manual d'instruccions de la sonda diferencial

Model: N1070A

1. Introducció

The PINTECH N1070A is a high-performance differential probe designed for safe and accurate measurement of high voltage signals with an oscilloscope. It features a bandwidth of up to 50MHz and can handle differential voltages up to 7kVp-p, making it suitable for various power electronics and industrial applications. This manual provides essential information for the proper setup, operation, and maintenance of your N1070A differential probe.

2. Informació de seguretat

ADVERTIMENT: This device measures high voltages. Failure to follow safety precautions can result in serious injury or death. Always adhere to the following safety guidelines:

  • Always connect the probe's ground lead to the oscilloscope's ground terminal before connecting the input leads to the circuit under test.
  • Ensure the probe is rated for the maximum voltage of the circuit being tested. The N1070A has a maximum differential test voltage of 7kVp-p.
  • No supereu el volum màxim d'entradatage ratings specified in the technical specifications.
  • Inspect the probe, leads, and accessories for any damage before each use. Do not use if any part is damaged.
  • Operate the probe only in dry conditions. Avoid exposure to moisture or extreme temperatures.
  • Use only the provided adapter or appropriate batteries for power.
  • Refer to your oscilloscope's manual for proper connection and operation procedures.
  • Servicing should only be performed by qualified personnel.

3. Contingut del paquet

Verifiqueu que tots els elements siguin presents al paquet:

  • N1070A Differential Probe x 1
  • Adaptador d'alimentació x 1
  • Probe Hook (Red) x 1
  • Probe Hook (Black) x 1
  • Manual d'instruccions x 1
PINTECH N1070A High Voltage Differential Probe and accessories

Figure 3.1: Contents of the N1070A package, including the differential probe unit, power adapter, and red and black probe hooks.

4. Configuració

  1. Font d'alimentació: The N1070A probe can be powered by either the included power adapter or internal batteries (not included).
    • To use the adapter, connect it to the probe's power input jack and then to a suitable power outlet.
    • To use batteries, open the battery compartment (refer to the diagram on the probe unit) and insert the required batteries, ensuring correct polarity.
  2. Connectar-se a l'oscil·loscopi: Connect the BNC cable from the N1070A probe's output to an input channel on your oscilloscope. Ensure the connection is secure.
  3. Select Attenuation: On the N1070A probe unit, select the desired attenuation ratio (100X or 1000X) using the switch. This setting determines the probe's voltage division ratio.
  4. Configuració de l'oscil·loscopi: Set your oscilloscope's input impedance to 1MΩ. Adjust the vertical scale (Volts/Div) on your oscilloscope according to the selected attenuation ratio. For example, if the probe is set to 100X, a 1V/Div setting on the oscilloscope will represent 100V/Div at the probe tips.

5. Instruccions de funcionament

  1. Encès: Turn on the N1070A probe using the power switch. The power indicator light should illuminate.
  2. Desplaçament zero: The probe features a built-in auto-zero function. Allow a few moments for the probe to stabilize after powering on.
  3. Connexió al circuit: Carefully connect the red and black input leads of the differential probe to the two points in the circuit where you wish to measure the voltage difference. Use the provided probe hooks for secure connection.
  4. Observe Waveform: Observe the waveform on your oscilloscope. Adjust the oscilloscope's time base and vertical scale as needed for a clear display.
  5. Protecció de sobrecàrrega: The probe includes overload protection. If the input voltage exceeds the probe's limits, the output may clip or distort. Disconnect immediately and verify the voltage gamma.
  6. Apagar: After use, disconnect the probe from the circuit under test, then turn off the probe and the oscilloscope.

6. Manteniment

  • Neteja: Clean the probe's exterior with a soft, dry cloth. Do not use abrasive cleaners or solvents. Ensure the probe is powered off and disconnected before cleaning.
  • Emmagatzematge: Store the probe in a clean, dry environment, away from direct sunlight and extreme temperatures.
  • Substitució de la bateria: If using batteries, replace them when the power indicator dims or the probe malfunctions. Ensure correct polarity during replacement.
  • Inspecció: Regularly inspect the probe body, cables, and connectors for any signs of wear or damage.

7. Solució De Problemes

  • Sense energia:
    • Comproveu si l'adaptador de corrent està connectat correctament i funciona.
    • If using batteries, ensure they are correctly inserted and not depleted.
  • No Signal on Oscilloscope:
    • Verify that the BNC cable is securely connected to both the probe and the oscilloscope.
    • Ensure the probe is powered on.
    • Check the oscilloscope's input channel settings (e.g., input impedance, vertical scale).
    • Confirm the probe leads are properly connected to the circuit under test.
  • Distorted or Clipped Waveform:
    • El volum d’entradatage may exceed the probe's maximum rating for the selected attenuation. Switch to a higher attenuation ratio (e.g., 1000X) or reduce the input voltage.
    • Check the oscilloscope's vertical scale setting; it might be too sensitive.

8. Especificacions

Table of technical specifications for PINTECH N1070A differential probe

Figure 8.1: Detailed technical specifications for the N1070A High Voltage Differential Probe.

ParàmetreValor
ModelN1070A
Ample de banda (-3 dB)100X: DC-40MHz; 1000X: DC-50MHz
Attenuation Rate1000X/100X
Precisió±1%
Temps de pujada≥7ns
Prova de voltage (Vp-p)100X: 700V; 1000X: 7kV
Entrada DC Voltage100X: 350V; 1000X: 3.5kV
Differential-mode voltage (Vp-p)100X: 350V; 1000X: 3.5kV
Valor RMS2450 V
Vol en mode comútagi RMS2450 V
Output (into 50kΩ load)<7V
Offset (típic)<± 5 mV
Input Impedance (Between two ends)20MΩ//0.5pF
Input Impedance (Single-ended to ground)10MΩ//1pF
Impedància d'entrada de l'oscil·loscopi1 MΩ
CMRR (60Hz)> 80 dB
CMRR (100kHz)> 50 dB
Noise RMS (100X)<100 mV
Noise RMS (1000X)<500 mV
Pes425 g / 0.94 lliures
Mida (LxWxH)187 x 73 x 41 mm
Temperatura de funcionament0-50°C
Humitat de funcionament20-75% HR

9. Garantia i Suport

PINTECH offers a 1 any de garantia for the N1070A differential probe from the date of purchase. This warranty covers defects in materials and workmanship under normal use.

For technical support, warranty claims, or any questions regarding the product, please contact PINTECH customer service. Refer to the contact information provided with your purchase documentation or visit the official PINTECH weblloc.

PINTECH is committed to providing professional advice and support to meet your measurement needs.

© 2023 PINTECH. All rights reserved.