HYTAIS TS380

Manual d'instruccions

Termòmetre infraroig sense contacte

Models: TS380 & TS600

Brand: HYTAIS

1. Introducció

This series of infrared thermometers is designed to measure the surface temperature of opaque objects. The optical components of the thermometer can sense the reflected and transmitted energy emitted by the object and focus the energy on the detector, making it suitable for quickly measuring the surface temperature of various high-temperature, toxic, or difficult-to-reach objects.

The thermometer consists of an optical system, a photoelectric sensor, a signal amplifier, a signal processing circuit, and an LCD display. The optical system converges the infrared energy radiated from the object's surface to the photoelectric sensor, which converts the energy into a corresponding electrical signal. This signal is then converted into a reading by the signal amplifier and signal processor circuit and displayed on the LCD.

2. Instruccions de seguretat

Avís:

  • To avoid potential harm to users, please read and follow the instructions below:
  • Do not point the laser at eyes or indirectly through reflective surfaces towards eyes when using this device.
  • The thermometer cannot measure through transparent surfaces such as glass/plastic, otherwise the measured value will be the surface temperature of the transparent object.
  • Steam/dust/smoke or other particles can create obstacles on the lens of the instrument, affecting the accuracy of the measurement.

Atenció:

  • To avoid damaging the thermometer or the measured equipment, please protect them from the following:
  • EMF (Electromagnetic Fields) generated by arc welding machines, induction heaters, etc.
  • Thermal shock (large or sudden changes in temperature caused by the environment, wait for 30 minutes before use to allow the thermometer to reach a stable state).
  • Do not place or bring the thermometer close to high-temperature objects.

3. Configuració

Instal·lació de la bateria

Open the battery door and install two 1.5V AAA batteries according to their polarities.

Battery compartment of the HYTAIS infrared thermometer
Figura 1: The battery compartment of the HYTAIS infrared thermometer, showing where to insert two AAA batteries.

Encès inicial

Pull the trigger to turn on the device. The device is ready for measurement once powered on.

Vídeo 1: Demonstration of battery installation and basic operation of the HYTAIS infrared thermometer, including trigger pull, function button usage, Celsius/Fahrenheit switch, and temperature measurements of various objects like a steamer and refrigerator compartments.

4. Instruccions de funcionament

Mesura ràpida

Aim the positioning laser at the surface of the object to be measured (if laser positioning is not needed, it can be turned off). Pull the trigger and the target temperature will be displayed on the LCD.

Hand holding HYTAIS infrared thermometer measuring temperature with laser pointer
Figura 2: A hand holding the yellow and black infrared thermometer, pointing it forward with a red laser dot visible. The display shows "34.3°C". This image highlights features like 0.5-second quick measurement, laser positioning, adjustable emissivity, high-low temperature alert, and Celsius/Fahrenheit switch.

Localització de punts freds o calents

To find a hot or cold spot, first aim the thermometer at the outer edge of the area to be measured. Then slowly move the thermometer up and down to scan the entire area or until the location of the hot or cold spot is determined.

HYTAIS infrared thermometer measuring a plant
Figura 3: A hand holding the infrared thermometer, measuring the temperature of a plant.
HYTAIS infrared thermometer measuring water temperature
Figura 4: A hand holding the infrared thermometer, measuring the temperature of water in a clear container.

Botons de funció

Labeled diagram of HYTAIS infrared thermometer components
Figura 5: Dos views of the infrared thermometer, one from the side and one from the front, with labels pointing to key components like the laser emission point, temperature sensing area, measurement trigger, battery compartment, LCD display with backlight, Celsius/Fahrenheit switch, and function switch.
  • Activador: When pulled, the temperature value is displayed, and a measurement symbol will appear. When the trigger is released, the LCD displays a 'HOLD' character and saves the current data. If there is no operation within a few seconds, the thermometer will automatically turn off.
  • Positioning Laser: After pulling the trigger and releasing it, press the '▲' button to turn on or off the positioning laser. The laser positioning symbol will appear or disappear on the LCD.
  • Conversió d'unitats de temperatura: After pulling the trigger and releasing it, press the '°C/°F' button to switch between Celsius and Fahrenheit temperature units.
  • Function Key (MODE): After pulling the trigger and releasing it, press the 'MODE' button to cycle through 'E-MAX-Hi-LOW' displayed at the bottom of the LCD.
    • E: Display the current emissivity value set.
    • MAX: Displays the maximum value of measured data.
    • Hi: Displays the current set maximum temperature value.
    • LOW: Displays the current set minimum temperature value.

Configuració d'emissivitat

After pulling the trigger and releasing it, press and hold the 'MODE' button. The LCD displays the current emissivity (e.g., 'E=0.95'). Use the '▲' or '▼' buttons to adjust the emissivity value (range 0.1-1.00).

Emissivity describes the characteristics of a material's radiated energy. Most organic materials and painted or oxidized surfaces have an emissivity of 0.95.

The preset emissivity of this series of temperature measurement instruments is 0.95. Measuring a shiny metal surface may result in inaccurate readings. To compensate for the readings, you can cover the surface to be measured with opaque adhesive paper or flat black paint (less than 148°C/300°F). Wait for a while to allow the adhesive paper or paint to reach the same temperature as the covered surface, and then measure the temperature of the adhesive paper or paint surface.

HYTAIS infrared thermometer display showing adjustable emissivity
Figura 6: A hand holding the infrared thermometer, showing the display with "34.3°C" and "34.0". Text overlay states "adjustable emissivity: 0.1-1.00" and "Adjustable emissivity for different material surfaces to achieve precise measurement".

Common Emissivity Values

Nom del materialEspecificacióEmissivitatNom del materialEspecificacióEmissivitat
PlomOxidat0.20-0.40Pell humana0.98
coureOxidat0.40-0.80Grafit0.98
courePolit0.02-0.05PlàsticTransparency > 0.5mm0.95
OrPolit0.01-0.10Cautxú0.95
FerroOxidat0.60-0.90Amiant0.95
AcerOxidat0.70-0.90formigó0.95
Guix0.80-0.90Sòl0.98
Asfalt0.95Estuc0.91
Ceràmica0.95Maó0.96
Fusta0.90-0.95Marbre0.94
pols de carbó vegetal0.96Tèxtil0.90
LacaAcabat mat0.94PaperDiversos Colors0.97
Carbon Rubber0.90Sorra0.90
Soap Bubbles0.92-0.95Fang0.85-0.92
Aigua0.93Grava0.95
Neu0.83-0.90Vidre0.96
Gel0.96-0.98Vaixella0.95

High/Low Temperature Alarms

Alarma d'alta temperatura: After pulling the trigger and releasing it, press and hold the 'MODE' button, then briefly press the 'MODE' button to select the high temperature alarm setting. The LCD defaults to 'Hi 380/600°C'. Use the '▲' or '▼' buttons to set the alarm temperature point. When the measured temperature exceeds the set temperature point, the indicator light above the display window will turn on.

Alarma de baixa temperatura: After pulling the trigger and releasing it, press and hold the 'MODE' button, then briefly press the 'MODE' button to select the low temperature alarm setting. The LCD defaults to 'LOW -50°C'. Use the '▲' or '▼' buttons to set the alarm temperature point. When the measured temperature is below the set temperature point, the indicator light above the display window will turn on.

HYTAIS infrared thermometer showing temperature warning limits
Figura 7: A hand holding the infrared thermometer. The display shows "34.3°C". Text overlay indicates "temperature warning" and "setting temperature upper and lower limits" with red and green boxes for "Upper limit" and "Lower limit".

5. Especificacions

Especificacions generals

CaracterísticaTS380TS600
Measurement Range (°C)-50 °C ~ 380 °C-50 °C ~ 600 °C
Measurement Range (°F)-58 ° F ~ 716 ° F-58 ° F ~ 1112 ° F
Precisió± (1.5% + 1.5)
Resolució0.1 °C / 0.1 °F
Emissivitat ajustable0.1 ~ 1.00.95 (fix)
Relació distància-punt (D:S)12:1
°C/°F Temperature Switch
Apagat automàtic
Retenció de dades
Llum de fons
Posicionament làser
PoderAAA*2 batteries
EstilPortàtil
ÚsIndustrial
CertificacióCE
Table of specifications for HYTAIS TS380 and TS600 models
Figura 8: A table detailing the functional parameters (specifications) for models TS380 and TS600, including measurement range, accuracy, resolution, adjustable emissivity, distance-to-spot ratio, temperature switch, auto power off, data hold, backlight, positioning laser, and power (AAA*2 batteries).

Relació distància-punt (D:S)

When using this device to measure temperature, it is important to consider the ratio between the distance and the size of the measurement area (referred to as the distance-to-spot ratio). As the distance between the device and the object being measured increases, the measurement area will also increase accordingly. (The distance-to-spot ratio of this instrument is 12:1). This instrument is equipped with a positioning laser for aiming at the surface of the object being measured.

Measurement Range: It is crucial to ensure that the target being measured is larger than the diameter of the light spot of the thermometer. The smaller the target is, the closer it should be to the target. For accurate measurement, ensure that the target being measured is at least twice the diameter of the light spot of the thermometer.

Diagram illustrating Distance-to-Spot ratio for infrared thermometer
Figura 9: A diagram illustrating the Distance-to-Spot (D:S) ratio of 12:1 for the thermometer, showing how the measurement area increases with distance. Text explains that the diameter of the object being measured is 1 unit, and it can be tested up to 12 times the diameter, recommending not to exceed 1 meter for testing.

6. Manteniment

Neteja de lents

Use clean compressed air to blow away debris, then use a camel hair brush to remove any residual tiny debris. Finally, use a damp cotton cloth to carefully wipe the surface.

Neteja de closques

Use a wet sponge or soft cloth with soap and water to clean.

Important Notes for Cleaning

  • Do not use any solvents to clean the plastic lens.
  • No submergiu el termòmetre en aigua.

7. Solució De Problemes

If the device is not functioning as expected, please ensure batteries are correctly installed and charged. Refer to the operating instructions for proper usage. If issues persist, contact customer support.

8. Consells d'usuari

  • The device is very simple to operate; just pull the trigger for one-key measurement.
  • Enjoy quick measurements in just 0.5 seconds.
  • This device is designed for industrial use and is not intended for measuring humans or animals.
  • Ensure correct emissivity setting for precise measurements on different materials.

9. Garantia i Suport

Política de garantia

Thank you for choosing our company's product. We provide our customers with excellent after-sales service to address any concerns you may have. The warranty terms are as follows:

  • El període de garantia d'aquest producte és d'un any.
  • During the warranty period, if a malfunction occurs under normal use conditions as specified in the user manual, free repairs will be provided with the original warranty card and purchase receipt. You can either send the product directly to our company or hand it over to the dealer.
  • Damages caused by improper use, natural wear and tear, unauthorized disassembly for self-repair, or accidents are not covered by the free warranty.
  • For cases not covered by the free warranty, our company can provide non-free repairs and charge accordingly.
  • The product must be repaired by our company's professional staff. If unauthorized disassembly or repair is performed, our company will not provide a warranty.

Assistència addicional

For additional information or support, please refer to the official user manual (PDF) available at: Manual d'usuari (PDF).

Documents relacionats - TS380

Preview HYTAIS TS380/TS600 Infrared Thermometer User Manual
User manual for HYTAIS TS380 and TS600 non-contact infrared thermometers, detailing features, safety instructions, operation, maintenance, and specifications for accurate surface temperature measurement.
Preview HYTAIS TS600 Digital Infrared Thermometer Gun User Manual
User manual for the HYTAIS TS600 digital infrared thermometer gun, detailing its features, operation, safety precautions, maintenance, and warranty information.
Preview HYTAIS TS-U7 Fully Automatic Recognition Multimeter User Manual
Comprehensive user manual for the HYTAIS TS-U7 Fully Automatic Recognition Multimeter. Learn about its features, specifications, operation, and safety guidelines for accurate electrical measurements.
Preview HYTAIS TS20A Pen-Type Digital Multimeter User Manual
User manual for the HYTAIS TS20A pen-type digital multimeter, detailing its features, operation, safety precautions, and specifications for measuring AC/DC voltage, resistència i continuïtat.