TOOLTOP HFS-20PRO

HFS-20PRO Handheld Nuclear Radiation Detector

Manual d'usuari

Model: HFS-20PRO | Brand: TOOLTOP

1. Introducció

The TOOLTOP HFS-20PRO is a small, highly sensitive radiation dose alarm detector designed for monitoring X, beta (β), and gamma (γ) rays. It utilizes a high-performance semiconductor Geiger counter with energy compensation, offering superior consistency, linearity, and reliability compared to traditional glass or metal counters. This design also mitigates the risk of air leakage failure. The device features a TFT 2.0 color screen for clear display and is easy to operate. It provides both sound and light alarms, with customizable alarm thresholds.

Davant view of the HFS-20PRO detector showing its display and key features.
Figure 1.1: HFS-20PRO Personal Radiation Dosimeter highlighting real-time measurement, dual alarm, and TFT 2.0 color screen.

2. Aplicacions

This detector is widely used in various environments requiring radiation dose monitoring and protection, including:

  • Family building materials radiation detection
  • Irradiation processing enterprises
  • Health and epidemic prevention
  • Radiation therapy facilities
  • Nuclear laboratories and nuclear power plants
  • Import and export commodity inspection
  • Indústria petroquímica
  • Geological surveys
  • Scrap steel inspection
  • Industrial non-destructive testing
Illustrations of various environments where the radiation detector can be used, such as family life, radiology departments, industrial settings, and food radiation detection.
Figure 2.1: Wide application range of the HFS-20PRO detector.

3. Safety Information and Dose Limits

Understanding radiation dose limits is crucial for safety. The HFS-20PRO helps monitor exposure to ensure these limits are not exceeded.

Human Cumulative Dose Limits:

Categoria Dose Limit
Radiation Industry Personnel (Average annual dose for 5 consecutive years) 20 mSv
Radiation Industry Personnel (Effective dose in any given year) 50 mSv
Radiation Industry Personnel (Annual equivalent dose of the lens of the eye) 150 mSv
Radiation Industry Personnel (Annual equivalent dose to extremities or skin) 500 mSv
Public (Annual cumulative dose) 1 mSv
Public (Effective dose for the year, if 5-year average does not exceed 1mSv) 5 mSv
Public (Annual equivalent dose of the lens of the eye) 15 mSv
Public (Annual equivalent dose to extremities or skin) 50 mSv

International Standard: Radiation industry personnel: 20mSv/year (10uSv/hour); Ordinary personnel: 1mSv/year (0.52uSv/hour).

Illustrations depicting potential health hazards of nuclear radiation, including cataract, lung problems, leukemia, alopecia, skin disease, and thyroid problems.
Figure 3.1: Potential hazards of nuclear radiation.

4. Contingut del paquet

Si us plau, reviseu el paquet amb cura per assegurar-vos que hi hagi tots els elements:

  • 1 x HFS-20PRO Nuclear Radiation Detector
  • 1 x cable USB tipus C
  • 1 x Corretja de canell
  • 1 x Operation Manual (this document)
Image showing the HFS-20PRO detector, its retail box, a USB Type-C cable, a wrist strap, and the operation manual.
Figure 4.1: Contents included in the HFS-20PRO package.

5. Producte acabatview

The HFS-20PRO detector is designed for ease of use and durability. Familiarize yourself with its components and controls:

Diagram labeling the parts of the HFS-20PRO detector: Light tips, display screen, Buzzer, On/off and selection key, Charging interface, and navigation buttons.
Figure 5.1: Labeled components of the HFS-20PRO detector.
  • Pantalla de visualització: TFT 2.0 color screen for real-time data display.
  • Light Tips: Indicator light (e.g., green for charging/full, alarm light).
  • Zumbador: Provides audible alarms and feedback.
  • Interfície de càrrega: USB Type-C port for charging the built-in battery.
  • Botons de control:
    • (Play/Pause/Confirm): Startup, shutdown, and function confirmation.
    • (Left/Decrease): Move cursor left or decrease numerical values.
    • (Right/Increase): Move cursor right or increase numerical values.
Close-up images showing anti-slip concave design, buzzer hole, comfort buttons, and Type-C charging interface.
Figure 5.2: Detailed features including anti-slip design, buzzer, and charging port.
Enrere view of the HFS-20PRO detector showing the 'Sensitive Area' label and product information.
Figura 5.3: posterior view indicating the sensitive area for radiation detection.

6. Especificacions

Característica Detall
Nom del producte HFS-20 Nuclear Radiation Detector
Model HFS-20PRO
Sensor Geiger-Muller counter tube (Energy compensation)
Ray Detection β (beta), γ (gamma), and X rays
Intervals de mesura 0.08uSv-50mSv (Dose Rate), 0.1uSv-9999mSv (Total Dose)
Range Response 33keV-3MeV
Precisió de la mesura -17% to +25% (based on 137 Cs γ)
Alarm Response Time (td) <10 s
Alarm Threshold Deviation -17% a +25%
Mostra TFT 2.0 color large screen
Bateria Bateria de liti de 750 mAh integrada
Temps de resistència 14 days (continuous measurement without particle sound) / 20 days (no particle sound on)
Interfície de càrrega Connector TYPE-C
Entorn de treball -20 °C a 45 °C (-4 °F a 113 °F)
Dimensions 89 * 34 * 19 mm (3.50 * 1.34 * 0.75 polzades)
Pes 57g (Item), 220g (Package)
Diagram showing the dimensions of the HFS-20PRO detector and a table of product parameters including sensor, measuring range, accuracy, endurance, charging interface, and alarm response time.
Figure 6.1: Product dimensions and key parameters.
Diagram illustrating the use of a GM counter sensor with its measuring range, accuracy, and measurable ray types.
Figure 6.2: GM counter sensor details.

7. Configuració i ús inicial

7.1 Càrrega del dispositiu

The HFS-20PRO comes with a built-in rechargeable lithium battery. Before first use, or if the device has been stored for a long time, ensure it is fully charged.

  1. Connecteu el cable USB tipus C proporcionat al port de càrrega del dispositiu.
  2. Connecteu l'altre extrem del cable a un adaptador de corrent USB estàndard (no inclòs) o a un port USB de l'ordinador.
  3. The green indicator light will be on during charging.
  4. The green light will turn off once the battery is fully charged.

Note: Always charge the device in time to maintain detection accuracy. If the particle sound is enabled, the battery endurance time may be shortened.

Diagram showing the built-in lithium battery and Type-C charging port, with details on endurance and charging indicator light.
Figure 7.1: Built-in lithium battery and Type-C charging.

8. Instruccions de funcionament

8.1 Encès/Apagat

  • Per encendre: Long press the "" button for 3 seconds.
  • Per apagar: Long press the "" button for 3 seconds.

Note: If the "" icon appears at the top of the screen (indicating a scheduled measurement is active), the device cannot be shut down. Enter the "Schedule" page to stop the timing first, then shut down.

8.2 Navigation and Menu Overview

Use the "" i "" buttons to navigate through menus and adjust values. The "" button confirms selections.

Image showing the main menu icons of the HFS-20PRO detector: Measure, Schedule, Records, Delete, Alarm, Unit, Sound, Clock, Sleep.
Figure 8.1: Main menu interface.

8.3 Radiation Measurement

From the main interface, click "Measure" to view real-time radiation data:

  • Taxa de dosi: Instantaneous radiation dose rate (e.g., µSv/h, µR/h).
  • Dose Rate AVG: Average dose rate.
  • CPM: Counts Per Minute.
  • Total Dose: Accumulated radiation dose.
Close-up of the TFT 2.0 color screen displaying dose rate, average dose rate, CPM, and total dose.
Figure 8.2: Real-time measurement display.

Mode de prova:

  • Lent: Default mode. Includes an internal filtering algorithm to avoid false alarms from external interference. Measurement delay is greater than 5 seconds.
  • Ràpid: For measuring transient X-rays or pulse radiation (e.g., from CT machines). Switch to this mode for quicker response.

8.4 Schedule (Timed Measurements)

This function allows you to set a delayed start and a specific measurement duration.

  1. Enter the "Schedule" menu.
  2. Set the start delay time (00-99 hours, minutes, seconds).
  3. Set the measurement time (00-999 hours).
  4. Click the "" button to start.

The device will count down the delay time. Once the delay countdown reaches zero, the instantaneous value, average value, and accumulated dose will be zeroed, and the device will begin counting radiation dose and cumulative dose rate for the set measurement time. At the end of the measurement time, the data is automatically stored in history.

8.5 Records (Historical Data)

The device automatically records data after each timed measurement. It can save up to 100 groups of historical records. When the memory is full, the oldest record will be overwritten by new data.

8.6 Suprimir dades

In this interface, you can delete current or historical data.

  • Delete Current Data: Removes the average dose rate and cumulative time from the radiation measurement page.
  • Delete History Data: Deletes all historical data records.

Note: If the "" icon is on the screen, data deletion cannot be performed. Stop any active scheduled measurements first.

8.7 Configuració d'alarma

The HFS-20PRO features sound and light dual alarms. You can set custom alarm thresholds.

  • Over-limit Alarm Setting Range:
    • Dose Rate: 0.2uSv/h - 999uSv/h (or 20µR/h - 99.9Mr/h)
    • Total Dose: 1uSv - 100000uSv (or 0.1Mr - 10000mR)

When the detected value exceeds the set alarm threshold, the detector will activate both sound and light alarms.

Diagram illustrating the sound and light dual alarm system of the detector.
Figure 8.3: Sound and light dual alarm in action.

8.8 Unit and Language Settings

In the "Unit" interface, you can switch between measurement units and display languages.

  • Unitats de mesura: Switch between "uSv" and "uR".
  • Idiomes: Canvia entre xinès i anglès.
Screenshots showing the Chinese and English language interfaces of the detector.
Figure 8.4: Chinese/English language switching.

8.9 Configuració de so

The "Sound" interface allows you to adjust alarm volume and enable/disable count sounds.

  • Volum de l'alarma: Set the desired volume level. Setting to '0' will turn off the alarm sound.
  • Count Sound: Toggle "ON" or "OFF" for the audible click sound during detection.

8.10 Configuració del rellotge

In this interface, you can set the device's internal clock (year-month-day and hour:minute:second).

8.11 Sleep Settings

Configure power-saving options in the "Sleep" interface.

  • Screen Sleep (min): Set the time after which the screen will turn off. Select "to keep screen constant light" for always-on display.
  • Auto Shutdown (hour): Set the time after which the device will automatically power off if inactive.
  • Brillantor de la pantalla: Adjust the screen brightness level (0-10).

8.12 Calibration Factor

The calibration factor can be adjusted for precise measurements. This is an advanced setting.

  1. Premsa "" to enter the delete surface.
  2. Double click "Calibration factor" at the bottom of the screen.
  3. A "Password:0000" prompt will appear.
  4. Click the right button "" to change the password to "0018" (for calibrating errors 200uSv/h~1uSv/h) or "0020" (for calibrating errors above 200uSv/h).
  5. Feu clic a "" to confirm.
  6. Use the left "" or right "" button to modify the calibration factor.
  7. Finally, click "" to save the modified parameter.

Note: By default, C1=1.00, C2=1.00. Calibration factor C1 * C2 = personal dose equivalent rate Hp(10) Reference value / instrument indication value.

9. Manteniment

  • Cura de la bateria: If the device is not used for a long time, ensure the battery is fully charged before storage. Charge promptly when the battery level is low to maintain detection accuracy.
  • Vibració: The sensor is sensitive to vibration. Avoid severe impacts or vibrations, as this may cause false alarms or damage the device.
  • Neteja: Feu servir un drap suau i sec per netejar el dispositiu. No utilitzeu productes de neteja abrasius ni dissolvents.
  • Emmagatzematge: Guardeu el dispositiu en un lloc fresc i sec, allunyat de la llum solar directa i de temperatures extremes.

10. Solució De Problemes

Problema Causa possible Solució
El dispositiu no s'encén Bateria baixa o mal funcionament del dispositiu. Charge the device fully. If it still doesn't power on, contact support.
Cannot shut down the device Scheduled measurement is active. Go to the "Schedule" menu and stop the active timing, then try to shut down.
False alarms or erratic readings Device subjected to severe vibration or external interference. Ensure the device is stable and in a calm environment. Switch to "Slow" test mode if needed.
"Device damaged" message Sensor fault detected. Contacteu amb l'assistència al client per obtenir assistència.
Data deletion not possible Scheduled measurement is active. Stop any active scheduled measurements first.

11. Consells d'usuari

  • For general environmental monitoring, the default 'Slow' test mode is usually sufficient and provides more stable readings.
  • If you suspect a sudden, short burst of radiation (e.g., near an X-ray machine), switch to 'Fast' test mode for a quicker response.
  • Regularly check the battery level, especially before critical measurements, to ensure uninterrupted operation.
  • Keep the device's sensitive area (back of the device) clear and unobstructed during measurements for optimal accuracy.
  • Utilize the timed measurement feature for consistent monitoring over specific periods, which can be helpful for trend analysis.

12. Garantia i Suport

Per obtenir informació sobre la garantia, assistència tècnica o consultes sobre el servei tècnic, consulteu la documentació proporcionada en el moment de la compra o poseu-vos en contacte amb el vostre distribuïdor. Conserveu el rebut de compra com a comprovant de compra.

You can also refer to the official TOOLTOP weblloc per obtenir recursos addicionals i informació de contacte.

Documents - TOOLTOP – HFS-20PRO