Corning 6795-620D

Corning PC-620D Stirring Hot Plate User Manual

Model: 6795-620D

1. Introducció

This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your Corning PC-620D Stirring Hot Plate. The PC-620D is a laboratory instrument designed for precise temperature control and stirring of various solutions, featuring a durable Pyroceram top and digital displays for both temperature and stirring speed.

Corning PC-620D Stirring Hot Plate with digital displays for temperature and RPM, and control knobs for stir and heat.

Figure 1: Corning PC-620D Stirring Hot Plate. This image shows the hot plate from an angled front view, highlighting the white Pyroceram top, the gray base, and the control panel with digital readouts for stirring speed (RPM) and temperature (Celsius), along with the 'Stir' and 'Heat' control knobs.

2. Informació de seguretat

Observeu sempre les següents precaucions de seguretat per evitar lesions i danys a l'instrument:

3. Què hi ha a la caixa

Upon unpacking, verify that the following item is present:

If any components are missing or damaged, contact your supplier immediately.

4. Configuració

  1. Col·locació: Place the hot plate on a stable, level, and heat-resistant surface away from flammable materials. Ensure there is sufficient clearance around the unit for air circulation.
  2. Connexió d'alimentació: Connect the power cord to the back of the hot plate and then plug the three-prong plug into a grounded 120V/60Hz electrical outlet.
  3. Inspecció inicial: Before first use, inspect the unit for any visible damage from shipping.

5. Funcionament

5.1 Controls i indicadors

5.2 Stirring Function

  1. Place a suitable stir bar into your vessel (e.g., beaker, flask) containing the solution.
  2. Place the vessel centrally on the Pyroceram top.
  3. Turn the STIR knob clockwise to the desired RPM. The digital display will show the set speed.
  4. Adjust the speed as needed for optimal mixing without splashing.
  5. To stop stirring, turn the STIR knob counter-clockwise to the "Off" position.

5.3 Funció d'escalfament

  1. Place your vessel containing the sample on the Pyroceram top.
  2. Turn the HEAT knob clockwise to the desired temperature. The digital display will show the set temperature in 5°C increments.
  3. The temperature display will blink until the set temperature is reached, then it will remain steady.
  4. Monitor the HOT TOP indicator. It will remain lit as long as the surface temperature is above 60°C, even after the heat is turned off.
  5. To turn off heating, turn the HEAT knob counter-clockwise to the "Off" position. Allow the unit to cool down naturally.

6. Manteniment i cura

7. Solució De Problemes

ProblemaCausa possibleSolució
La unitat no s'encén.No power to outlet; power cord disconnected or damaged.Check power outlet; ensure power cord is securely connected and undamaged.
No stirring action.Stir knob is off; stir bar not present or not coupled.Turn stir knob to desired RPM; ensure stir bar is in vessel and centered.
Sense calefacció.Heat knob is off; temperature set too low.Turn heat knob to desired temperature; increase set temperature.
Temperature display blinking.Unit has not reached set temperature.This is normal operation. Wait for the unit to reach the set temperature.

8. Especificacions

ParàmetreValor
Número de model6795-620D
Material de la placa superiorPyroceram Glass-Ceramic
Dimensions de la placa superior10 x 10 polzades (25.4 x 25.4 cm)
Interval de temperaturaDe 5 °C a 550 °C (41 °F a 1022 °F)
Increments de temperatura5 °C
Rang de velocitat d'agitació60 a 1150 RPM
Requisits elèctrics120V/60Hz
Dimensions totals (L x A x A)19 x 18 x 8 polzades
Pes11.46 lliures
CertificacionsUL, cUL Approved

9. Garantia i Suport

The Corning PC-620D Stirring Hot Plate comes with a 2 any de garantia from the date of purchase, covering defects in materials and workmanship under normal use. For warranty claims, technical support, or service inquiries, please contact Corning customer service or your authorized distributor.

Please have your model number (6795-620D) and purchase information ready when contacting support.

Documents relacionats - 6795-620D

Preview Manual d'instruccions de plaques calefactores, agitadors i agitadors/plaques calefactores Corning
Manual d'instruccions per a plaques calefactores, agitadors i agitadors/plaques calefactores amb pantalles digitals Corning i el controlador de temperatura 6795PR. Inclou instruccions d'ús, informació de seguretat, especificacions tècniques i detalls de la garantia.
Preview Manual d'instruccions de plaques calefactores, agitadors i agitadors/plaques calefactores Corning (120 V)
Manual d'instruccions complet per a les plaques calefactores, agitadors i agitadors/plaques calefactores de 120 V de Corning (sèries PC-200, PC-400, PC-600). Cobreix informació general, funcionament, manteniment, especificacions i seguretat.
Preview Manual d'usuari del sistema Corning Everon 6000 MIMO: configuracions, LED i especificacions de RF
Manual d'usuari detallat per als sistemes MIMO de la sèrie Everon 6000 de Corning Optical Communications. Cobreix configuracions 2x2, 2TO1 i 4TO1 amb components dLRU, DEU, DCU, RIU, BTS, dMRU i dHRU. Inclou definicions d'estat dels LED, versions de llançament i avisos d'exposició crítica a RF per a una instal·lació professional.
Preview Manual d'usuari de Corning MRU i HRU: Guia d'instal·lació i funcionament
Aquest manual d'usuari proporciona instruccions completes per a la instal·lació, configuració i funcionament dels sistemes MRU (Digital Medium-power Remote Unit) i HRU (High Power Remote Unit) de Corning, components essencials per a la infraestructura de xarxa sense fil.
Preview Corning Permeable Supports Selection Guide: Transwell and Falcon Cell Culture Inserts
A comprehensive guide from Corning and Fisher Scientific to select the optimal permeable supports, including Transwell and Falcon cell culture inserts, based on membrane type, pore size, format, and surface treatment for various research applications.
Preview Corning Matribot Bioprinter Instruction Manual
Comprehensive instruction manual for the Corning Matribot Bioprinter, detailing setup, operation, maintenance, and troubleshooting for advanced bioprinting applications in tissue engineering and 3D cell culture.