Corning PC-620D

Corning PC-620D Digital Stirring Hot Plate Kit

Manual d'usuari

1. Introducció

The Corning 6795-620KIT PC-620D Digital Stirring Hot Plate Kit is designed for precise heating and stirring of liquid solutions in laboratory environments. Featuring a durable 10 x 10 inch Pyroceram top plate, this unit offers fast heating, chemical resistance, and scratch resistance. Dual LED readouts provide clear display of temperature and speed settings, ensuring accurate control for various scientific applications.

This manual provides essential information for the safe and effective operation, setup, and maintenance of your Corning PC-620D Digital Stirring Hot Plate Kit.

2. Informació de seguretat

Observeu sempre les següents precaucions de seguretat per evitar lesions o danys a l'equip:

  • Seguretat elèctrica: Ensure the unit is connected to a properly grounded electrical outlet. Do not operate with damaged cords or plugs. Disconnect power before cleaning or servicing.
  • Risc de cremades: The hot plate surface can reach temperatures up to 550°C. Always assume the surface is hot, even if the hot top indicator light is off. Use appropriate personal protective equipment (PPE), such as heat-resistant gloves.
  • Seguretat química: Exercise caution when heating or stirring flammable, corrosive, or hazardous chemicals. Ensure proper ventilation and containment.
  • Estabilitat: Place the hot plate on a stable, level, and heat-resistant surface. Ensure glassware and support rods are securely positioned.
  • Sobreescalfament: Do not operate the unit unattended, especially when heating volatile substances. Monitor temperature settings carefully.
  • Ventilació: Ensure adequate ventilation to dissipate heat and any fumes generated during operation.

3. Què hi ha a la caixa

En desembalar, comproveu que hi hagi tots els components:

  • Corning PC-620D Digital Stirring Hot Plate
  • Controlador de temperatura extern
  • (2) 9-inch Support Rods
  • Stir Bar Retriever
  • Manual d'instruccions (aquest document)

4. Configuració

Follow these steps for initial setup of your hot plate:

  1. Col·locació: Select a stable, level, and heat-resistant workbench or surface away from flammable materials and direct sunlight. Ensure adequate clearance around the unit for ventilation.
  2. Support Rods: Insert the two 9-inch support rods into the designated threaded holes at the rear of the hot plate. These can be used to hold the external temperature controller or other accessories.
  3. Controlador de temperatura extern: Connect the external temperature controller probe to the appropriate port on the hot plate. Secure the controller using one of the support rods if desired.
  4. Connexió d'alimentació: Ensure the hot plate's power switch is in the 'OFF' position. Plug the power cord into a grounded 120V/60Hz electrical outlet.
Corning PC-620D Digital Stirring Hot Plate Kit

Figure 1: Corning PC-620D Digital Stirring Hot Plate Kit. This image shows the main unit with its control knobs and digital displays, and the Pyroceram top plate.

5. Instruccions de funcionament

The PC-620D features intuitive controls for both heating and stirring functions.

5.1. Funció de calefacció

  1. Encès: Turn the temperature control knob from 'OFF' to the desired temperature setting. The unit will power on, and the temperature LED readout will illuminate.
  2. Temperatura establerta: Adjust the temperature control knob to set the desired temperature. The temperature can be set from ambient +5°C up to 550°C in 5-degree increments. The dual LED display will show the set temperature.
  3. Monitorització de la temperatura: The hot plate will begin heating. The external temperature controller, if connected, helps ensure regulation to ±2°C.
  4. Hot Top Indicator: A 'HOT TOP' light will illuminate when the surface temperature is above 60°C, even if the power is off. This is a safety feature.

5.2. Stirring Function

  1. Place Stir Bar: Carefully place a stir bar (sold separately) into the vessel containing the liquid to be stirred.
  2. Encès: Turn the stirring speed control knob from 'OFF' to the desired speed setting. The unit will power on, and the speed LED readout will illuminate.
  3. Establir velocitat: Adjust the stirring speed control knob to set the desired RPM. The speed range is 60 to 1,150 rpm. Speed settability varies: 5 rpm at 60-100 rpm; 10 rpm at 100-200 rpm; 20 rpm at 200-400 rpm; 50 rpm at 400-1,150 rpm.
  4. Slow Start: The stirring control slowly increases speed for improved safety and enhanced coupling with the stir bar, especially for viscous solutions.

6. Manteniment

Proper maintenance ensures the longevity and optimal performance of your hot plate.

  • Neteja: Always disconnect the unit from power and allow it to cool completely before cleaning. Wipe the Pyroceram top plate with a damp cloth and mild detergent. For stubborn stains, a non-abrasive cleaner suitable for glass-ceramic surfaces can be used. Avoid abrasive pads or harsh chemicals.
  • Vessaments: Immediately clean any spills on the hot plate surface to prevent residue buildup or damage.
  • Atenció general: Keep the unit free from dust and debris. Do not immerse the unit in water. Inspect the power cord for any signs of damage regularly.

7. Solució De Problemes

If you encounter issues with your hot plate, refer to the following common troubleshooting steps:

  • La unitat no s'encén:
    • Check if the power cord is securely plugged into a live electrical outlet.
    • Ensure the temperature or stirring speed knob is turned from the 'OFF' position.
    • Verifiqueu que no s'hagi saltat l'interruptor o el fusible de la presa de corrent.
  • Hot Plate Not Heating:
    • Confirm the temperature setting is above ambient temperature.
    • Ensure the external temperature controller is properly connected and functioning if in use.
  • Stirring Not Functioning:
    • Ensure the stirring speed knob is turned from the 'OFF' position and set to a desired RPM.
    • Verify that a suitable stir bar is placed correctly in the vessel and is not stuck.
    • Check if the liquid viscosity is too high for the selected speed or stir bar.
  • Lectura de temperatura inexacta:
    • Ensure the external temperature controller probe is correctly positioned within the solution.
    • Verify the probe is clean and undamaged.

If problems persist after attempting these steps, contact Corning customer support for further assistance.

8. Especificacions

CaracterísticaEspecificació
Número de modelPC-620D (Kit: 6795-620KIT, Item Model: S5660-7K)
Material de la placa superiorGlass-ceramic (Pyroceram)
Top Plate Dimensions (W x D)10 x 10 polzades (25.4 cm x 25.4 cm)
Interval de temperaturaAmbient +5 °C a 550 °C
Precisió de la temperatura±2°C (with external temperature controller)
Interval de velocitat60 a 1,150 rpm
Speed Settability5 rpm (60-100 rpm); 10 rpm (100-200 rpm); 20 rpm (200-400 rpm); 50 rpm (400-1,150 rpm)
Poder120V/60Hz
Pes5.2 kg (11.46 lliures)
Dimensions totals (Alt x A x P)4.625 x 11 x 15.375 polzades (11.75 cm x 28 cm x 39.05 cm)

Documents relacionats - PC-620D

Preview Manual d'instruccions de plaques calefactores, agitadors i agitadors/plaques calefactores Corning
Manual d'instruccions per a plaques calefactores, agitadors i agitadors/plaques calefactores amb pantalles digitals Corning i el controlador de temperatura 6795PR. Inclou instruccions d'ús, informació de seguretat, especificacions tècniques i detalls de la garantia.
Preview Manual d'instruccions de plaques calefactores, agitadors i agitadors/plaques calefactores Corning (120 V)
Manual d'instruccions complet per a les plaques calefactores, agitadors i agitadors/plaques calefactores de 120 V de Corning (sèries PC-200, PC-400, PC-600). Cobreix informació general, funcionament, manteniment, especificacions i seguretat.
Preview Manual d'usuari del sistema Corning Everon 6000 MIMO: configuracions, LED i especificacions de RF
Manual d'usuari detallat per als sistemes MIMO de la sèrie Everon 6000 de Corning Optical Communications. Cobreix configuracions 2x2, 2TO1 i 4TO1 amb components dLRU, DEU, DCU, RIU, BTS, dMRU i dHRU. Inclou definicions d'estat dels LED, versions de llançament i avisos d'exposició crítica a RF per a una instal·lació professional.
Preview Manual d'usuari de Corning MRU i HRU: Guia d'instal·lació i funcionament
Aquest manual d'usuari proporciona instruccions completes per a la instal·lació, configuració i funcionament dels sistemes MRU (Digital Medium-power Remote Unit) i HRU (High Power Remote Unit) de Corning, components essencials per a la infraestructura de xarxa sense fil.
Preview Corning Permeable Supports Selection Guide: Transwell and Falcon Cell Culture Inserts
A comprehensive guide from Corning and Fisher Scientific to select the optimal permeable supports, including Transwell and Falcon cell culture inserts, based on membrane type, pore size, format, and surface treatment for various research applications.
Preview Corning Matribot Bioprinter Instruction Manual
Comprehensive instruction manual for the Corning Matribot Bioprinter, detailing setup, operation, maintenance, and troubleshooting for advanced bioprinting applications in tissue engineering and 3D cell culture.