1. Introducció
The Extech PH110 Waterproof ExStik pH Meter is designed for accurate pH measurements in various substances. This device features a refillable pH electrode, allowing for extended use and consistent accuracy by replenishing the reference solution. Its robust, waterproof design ensures reliability in diverse environments.
2. Característiques clau
- Refillable pH electrode for extended service life.
- Analog bargraph display on LCD for trend indication.
- Waterproof construction to IP57 standard.
- Automatic 1, 2, or 3 point calibration with buffer solution recognition.
- Flat Surface Electrode capable of measuring pH in liquids, semi-solids, and solids.
3. Contingut del paquet
The Extech PH110 Waterproof ExStik pH Meter package includes:
- Extech PH110 pH Meter with refillable pH electrode
- 1.5mL bottle of KCI (Potassium Chloride) solution
- Two reference junctions
- Two electrode removal tools
- Tapa protectora
- Sampla copa
- Four 3V CR-2032 batteries (pre-installed or included separately)
- 48-inch (1.2m) neck strap
4. Configuració
4.1 Instal·lació de la bateria
- Ensure the meter is off.
- Localitzeu el compartiment de la bateria.
- Insert four 3V CR-2032 Lithium Metal batteries, observing correct polarity.
- Tanqueu de manera segura el compartiment de la bateria.
4.2 Preparació dels elèctrodes
- Traieu la tapa protectora de l'elèctrode.
- Rinse the electrode tip with distilled water.
- If the electrode appears dry, soak it in electrode storage solution or pH 7 buffer for at least 30 minutes before first use.
5. Funcionament
5.1 Calibració
Calibration is essential for accurate measurements. The PH110 supports 1, 2, or 3 point calibration and automatically recognizes buffer solutions.
- Engegueu el mesurador prement el botó ON/OFF botó.
- Prepare fresh pH buffer solutions (e.g., pH 4, 7, and 10). These are sold separately.
- Manteniu premut el botó CAL fins que aparegui «CAL» a la pantalla.
- Immerse the electrode into the first buffer solution (e.g., pH 7). The meter will automatically recognize the buffer and calibrate. Wait for the reading to stabilize.
- For 2 or 3 point calibration, rinse the electrode with distilled water and immerse it into the next buffer solution (e.g., pH 4, then pH 10). The meter will continue to calibrate automatically.
- Ensure the display shows 'CAL' during the calibration process.
5.2 Mesurament
The flat surface electrode allows for versatile measurements.
- After calibration, rinse the electrode with distilled water.
- Submergiu l'elèctrode a la sample to be measured. For liquids, ensure the electrode tip is fully submerged. For semi-solids or solids, ensure good contact with the flat surface electrode.
- Allow the reading to stabilize on the display. The analog bargraph will indicate trends in acidity or alkalinity.
- To switch between pH and temperature display modes, press the MODE botó.
- After measurement, rinse the electrode thoroughly with distilled water and replace the protective cap.

Figura 1: The Extech PH110 pH meter, a handheld device with a digital display showing pH and temperature readings. It features three control buttons: MODE, CAL, and ON/OFF.
6. Manteniment
6.1 Electrode Refilling
The refillable electrode design allows for prolonged accuracy and lifespan.
- When the reference solution depletes or accuracy declines, use the provided removal tools to access the reference junctions.
- Carefully refill the electrode with the 1.5mL KCI solution.
- Ensure the reference junctions are properly re-seated after refilling.

Figura 2: Un primer pla view of hands refilling the Extech PH110 pH meter's electrode module with a specialized electrode filling solution from a small bottle. This process ensures the meter's continued accuracy and extends electrode life.
6.2 Neteja i emmagatzematge
- Always rinse the electrode with distilled water after each use to prevent contamination.
- The flat surface electrode is designed for easy cleaning. Gently wipe with a soft cloth if necessary.
- Store the electrode with its protective cap filled with electrode storage solution (or pH 7 buffer if storage solution is unavailable) to keep the sensor hydrated. Do not store dry.
7. Solució De Problemes
If the meter exhibits inaccurate readings or operational issues, consider the following:
- Lectures inexactes: Recalibrate the meter using fresh buffer solutions. Ensure the electrode is clean and properly hydrated. Check if the reference solution in the electrode needs refilling.
- Sense problemes de pantalla/alimentació: Verify that the batteries are correctly installed and have sufficient charge. Replace batteries if necessary.
- Resposta lenta: The electrode might be dry or contaminated. Clean and rehydrate the electrode, then recalibrate.
8. Especificacions
| Especificació | Detall |
|---|---|
| Fabricant | Extech |
| Número de part | PH110 |
| Pes de l'article | 3.87 unces |
| Dimensions del producte | 6.8 x 1.6 x 1.4 polzades |
| Número de model de l'article | PH110 |
| Bateries | 4 Lithium Metal batteries (CR-2032) required (included) |
| Font d'alimentació | Funciona amb bateria |
| Classificació impermeable | Estàndard IP57 |
| Calibració | 1, 2, or 3 point automatic buffer recognition |
| Tipus d'elèctrode | Refillable Flat Surface Electrode |
9. Informació de la garantia
Extech provides a full one-year warranty on parts and workmanship for the PH110 pH Meter, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modifications.
10. Suport
For further assistance, technical support, or to purchase accessories and replacement parts, please refer to the official Extech Instruments weblloc web o poseu-vos en contacte amb el seu departament d'atenció al client. La informació de contacte normalment es pot trobar a l'embalatge del fabricant o weblloc.