Introducció
The Hanna Instruments HI98304 DIST4 is a rugged, reliable, and pocket-sized tester designed for quick and accurate measurements of Electrical Conductivity (EC) and Total Dissolved Solids (TDS). This device is suitable for various applications including drinking water analysis, water conditioning, reverse osmosis systems, cooling towers, wastewater treatment, laboratories, agriculture, aquaculture, aquariums, hydroponics, and the printing industry. Its amperometric graphite electrode ensures consistent and dependable results by resisting oxidation.

This image displays the Hanna Instruments HI98304 DIST4 EC and TDS Tester. The device is light blue with a digital display showing '0.00 mS'. The model name 'DiST 4' and 'by HANNA' are printed on the body. The protective cap is shown removed, revealing the electrode at the bottom.
Configuració
1. Instal·lació de la bateria
The HI98304 tester requires one CR2 battery for operation. To install or replace the battery:
- Localitzeu el compartiment de la bateria al provador.
- Obriu la coberta del compartiment.
- Inseriu la bateria CR2, assegurant la polaritat correcta.
- Tanqueu de manera segura la tapa del compartiment de la bateria.
2. Calibració inicial
For accurate measurements, the tester requires a one-point manual calibration. It is recommended to perform calibration before first use and periodically thereafter.
- Ensure the electrode is clean and dry.
- Submerge the electrode tip into the appropriate calibration solution (HI70039 or HI70030).
- Locate the trimmer on the side of the tester.
- Using a small screwdriver, carefully adjust the trimmer until the display reads the value of the calibration solution.
- Rinse the electrode with distilled water after calibration.
Instruccions de funcionament
1. Prendre Mides
To obtain an EC or TDS reading:
- Traieu la tapa protectora de l'elèctrode.
- Enceneu el provador amb el botó d'engegada.
- Submergiu la punta de l'elèctrode a la sample solution, ensuring the liquid level is above the sensing portion of the electrode.
- Gently stir the tester to remove any air bubbles and allow the reading to stabilize.
- Read the value displayed on the LCD screen. The unit of measurement (mS/cm) will be indicated.
- Després de la mesura, esbandiu l'elèctrode amb aigua destil·lada i torneu a col·locar la tapa protectora.
2. Comprensió de lectures
The HI98304 displays readings in mS/cm (milliSiemens per centimeter) for Electrical Conductivity. The device automatically compensates for temperature variations between 0 to 50 ℃ (32 to 122 ℉).
Manteniment
1. Neteja d'elèctrodes
Regular cleaning of the electrode is crucial for maintaining accuracy and extending the life of your tester.
- After each use, rinse the electrode thoroughly with distilled or deionized water.
- If the electrode appears dirty or readings become erratic, clean it using a mild cleaning solution specifically designed for EC/TDS probes. Avoid abrasive materials.
- Never touch the graphite pins of the electrode with your fingers, as oils and residues can affect performance.
2. Emmagatzematge
When not in use, ensure the protective cap is securely placed over the electrode. Store the tester in a cool, dry place away from direct sunlight and extreme temperatures.
Resolució de problemes
If you encounter issues with your HI98304 tester, consider the following:
- Inaccurate or Erratic Readings:
- Ensure the electrode is clean. Refer to the Electrode Cleaning section.
- Perform a calibration. Refer to the Initial Calibration section.
- Check battery level. Replace the battery if low.
- Sense pantalla:
- Verify the battery is correctly installed and has sufficient charge. Replace if necessary.
- Assegureu-vos que el botó d'engegada estigui completament premut.
- Temps de resposta lent:
- Netegeu l'elèctrode.
- Allow sufficient time for the reading to stabilize in the sampla solució.
Especificacions
| Paràmetre | Valor |
|---|---|
| Interval | 19.99 mS / cm |
| Resolució | 0.01 mS / cm |
| Precisió | +/-2% FS |
| Factor TDS | 0.5 |
| Solució de calibratge | HI70039 or HI70030 |
| Calibració | Manual, 1 punt |
| Sonda | HI73304 |
| Compensació de temperatura | Automatic, 0 to 50 ℃ (32 to 122 ℉) |
| Tipus/Durada de bateria | 1 CR2 battery / Approximately 200 hours of continuous use |
| Medi ambient | 0 to 50 ℃ (32 to 122 ℉); RH max 95% |
| Pes | 3.4 oz (96.4 g) |
| Dimensions | 1-19/32" L x 57/64" W x 6-57/64" H (approx. 4.05 cm L x 2.26 cm W x 17.4 cm H) |
Informació de la garantia
Hanna Instruments products are warranted against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The specific warranty period may vary by product and region. Please retain your proof of purchase for warranty claims. For detailed warranty terms and conditions, please refer to the official Hanna Instruments weblloc o poseu-vos en contacte amb el seu servei d'atenció al client.
Suport
For technical assistance, troubleshooting beyond this manual, or inquiries regarding your Hanna Instruments HI98304 DIST4 tester, please contact Hanna Instruments customer support directly. You can find contact information on the official Hanna Instruments weblloc web o a través del vostre distribuïdor local.
Hanna Instruments Oficial Weblloc: www.hannainst.com