REED Instruments TM-8811-NIST

REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge User Manual

Model: TM-8811-NIST

1. Introducció

This manual provides comprehensive instructions for the safe and effective operation, maintenance, and troubleshooting of the REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge. This device is designed for precise measurement of material thickness using ultrasonic technology. Please read this manual thoroughly before using the instrument.

2. Informació de seguretat

Observe the following safety precautions to prevent injury and avoid damage to the instrument:

3. Producte acabatview

The REED Instruments TM-8811-NIST is an ultrasonic thickness gauge designed for quick and accurate measurements. It features a large LCD display and intuitive controls.

REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge with probes and a calibration certificate.

Figura 1: REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge and a sample calibration certificate.

This image displays the REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge, a blue handheld device with a digital display showing "0.000". Two probes are connected to the top of the device via cables. Below the display are several control buttons: a red power button (circle with a vertical line), "SELECT", "CAL", "VEL", "mm/inch", and up/down arrow buttons. The text "Ultrasonic Thickness Gauge" and "COUPLING" are visible on the device face. Partially visible in the background is a "CALIBRATION CERTIFICATE" document, indicating the importance of calibration for such instruments.

Components clau:

4. Configuració

  1. Instal·lació de la bateria: Open the battery compartment (typically on the back) and insert the required batteries, observing correct polarity. Close the compartment securely.
  2. Connexió de la sonda: Connect the ultrasonic probe(s) to the connectors at the top of the gauge. Ensure a firm connection.
  3. Encès: Premeu el botó Power button to turn on the device. The display will illuminate.

5. Instruccions de funcionament

5.1 Mesura bàsica

  1. Prepara la superfície: Ensure the surface of the material to be measured is clean, smooth, and free of rust, paint, or debris.
  2. Apply Couplant: Apply a small amount of ultrasonic couplant (e.g., glycerin, gel) to the surface where the measurement will be taken. This ensures proper sound transmission.
  3. Place Probe: Press the probe firmly and flatly against the prepared surface. The "COUPLING" indicator on the display should appear, confirming good contact.
  4. Llegir mesura: The thickness reading will be displayed on the LCD. If the reading is unstable, adjust the probe position or reapply couplant.
  5. Eliminar la sonda: Lift the probe from the surface. The reading will typically hold or return to zero.

5.2 Unit Selection (mm/inch)

Premeu el botó mm/polzada button to toggle between millimeters (mm) and inches (inch) for the displayed thickness measurement.

5.3 Sound Velocity Adjustment

The accuracy of ultrasonic thickness measurement depends on the correct sound velocity setting for the material being tested. Refer to a material velocity chart for common materials.

  1. Premeu el botó VEL button. The current sound velocity value will be displayed.
  2. Utilitza el Up i Avall arrow buttons to adjust the velocity to the desired value for your material.
  3. Premeu VEL de nou o SELECCIONA to confirm and exit velocity adjustment mode.

6. Calibració

The TM-8811-NIST gauge is factory calibrated. However, periodic calibration checks and adjustments are crucial for maintaining measurement accuracy, especially when measuring different materials or after prolonged use. The device may feature a built-in calibration block or require external reference standards.

6.1 Calibració zero

Zero calibration compensates for the probe delay line. This is typically performed by placing the probe on a known thickness standard or by coupling the probe to itself (if applicable to the probe type).

  1. Ensure the probe is clean and free of couplant.
  2. Premeu el botó CAL button. Follow the on-screen prompts.
  3. Place the probe on the zero calibration block or perform the specified zeroing procedure.
  4. Confirmeu el calibratge.

6.2 Calibració de dos punts

For enhanced accuracy, especially over a range of thicknesses, a two-point calibration using two known thickness standards (e.g., thin and thick) is recommended.

  1. Premeu el botó CAL button and select the two-point calibration option.
  2. Measure the first known thickness standard and input its value using the arrow buttons.
  3. Measure the second known thickness standard and input its value.
  4. Confirmeu el calibratge.

Always refer to the specific instructions in the device's internal menu for detailed calibration procedures.

7. Manteniment

8. Solució De Problemes

ProblemaCausa possibleSolució
No hi ha pantalla/El dispositiu no s'encénPiles gastades o instal·lades incorrectament.Comproveu la polaritat de la bateria; substituïu les piles.
Unstable or "0.000" readingPoor probe coupling; incorrect sound velocity; damaged probe.Reapply couplant, ensure firm probe contact; adjust sound velocity; inspect/replace probe.
"COUPLING" indicator not appearingInsufficient couplant; rough surface; probe not flat.Apply more couplant; prepare surface; ensure probe is flat.
Lectures inexactesIncorrect sound velocity; calibration required; surface conditions.Verify sound velocity setting; perform calibration; ensure proper surface preparation.

9. Especificacions

10. Garantia i Suport

REED Instruments products are designed for reliability and performance. For information regarding warranty coverage, technical support, or service, please refer to the official REED Instruments weblloc web o poseu-vos en contacte amb el seu departament d'atenció al client. Conserveu el rebut de compra com a comprovant de compra per a reclamacions de garantia.

Fabricant: Instruments REED

Data de primera disponibilitat: 10 de juliol de 2017

Documents relacionats - TM-8811-NIST

Preview Manual d'instruccions de la càmera d'imatge tèrmica REED R2160
Manual d'instruccions complet per a la càmera d'imatges tèrmiques REED R2160, que cobreix les característiques, les especificacions, el funcionament, la seguretat i l'assistència del producte.
Preview REED R1610 Thermo-Hygrometer Instruction Manual
This manual provides comprehensive instructions for the REED R1610 Thermo-Hygrometer, a Bluetooth® Smart Series instrument designed for accurate ambient temperature and humidity measurements. It covers features, specifications, setup, usage, care, and compliance information for safe and reliable operation.
Preview Manual d'instruccions del termòmetre de termopar REED R1640
Manual d'instruccions complet per al termòmetre de termopar REED R1640, amb connectivitat Bluetooth Smart Series. Cobreix les característiques del producte, les especificacions, els procediments d'operació, les directrius de seguretat i la informació d'assistència.
Preview Manual d'instruccions del mesurador de distància làser REED R8005
Manual d'instruccions complet per al mesurador de distància làser REED R8005, que cobreix les característiques, les especificacions, les instruccions de funcionament, els modes de mesura, la resolució de problemes i la informació sobre la garantia.
Preview Manual d'instruccions del mesurador de conductivitat/TDS/salinitat REED R3530
Manual d'instruccions del mesurador de conductivitat/TDS/salinitat REED R3530, que cobreix les característiques, les especificacions, el funcionament, la calibració, el manteniment i l'assistència tècnica del producte.
Preview Manual d'usuari i característiques del cronòmetre digital REED SW600
Guia completa del cronòmetre digital REED SW600, que cobreix les seves característiques, les funcions del cronòmetre, la configuració de l'alarma i la funció de repetició de l'alarma, la configuració de la data i l'hora i les instruccions d'ús.