AmScope T720

Manual d'usuari del microscopi compost trinocular de laboratori Kohler AmScope T720 40X-1000X Plan Infinity

Introducció

This manual provides detailed instructions for the proper setup, operation, and maintenance of your AmScope T720 Trinocular Compound Microscope. Please read this manual thoroughly before using the instrument to ensure optimal performance and longevity.

Informació de seguretat

Contingut del paquet

Verifiqueu que tots els elements que s'enumeren a continuació siguin presents al vostre paquet:

Producte acabatview

The AmScope T720 is a professional trinocular compound microscope designed for advanced laboratory and research applications. It features a Plan Infinity optical system, providing crisp and clear images across the entire field of view.

Davant-esquerra view of the AmScope T720 Trinocular Compound Microscope

Davant-esquerra view of the AmScope T720 Trinocular Compound Microscope, showcasing the trinocular head, eyepieces, revolving nosepiece with objectives, mechanical stage, and base with illumination controls.

Característiques principals:

Another perspective of the AmScope T720 microscope

Another perspective of the AmScope T720 microscope, showing the overall structure and the placement of the illumination and stage control mechanisms.

Configuració

1. Desembalatge i col·locació:

Carefully remove all components from the packaging. Place the microscope on a stable, level surface away from direct sunlight, heat sources, and vibrations.

2. Head Assembly:

Loosen the set screw on the top of the microscope body. Gently insert the trinocular head into the opening, ensuring it is fully seated. Tighten the set screw to secure the head. The head can be rotated 360 degrees for comfortable viewing.

3. Eyepiece Installation:

Remove the protective caps from the eyepiece tubes on the trinocular head. Insert the two 10X widefield eyepieces into the tubes. If dust caps are present, remove them.

4. Objective Installation:

Rotate the revolving nosepiece until the lowest power objective (4X) is positioned over the stage opening. Screw the 4X, 10X, 40X, and 100X Plan Achromatic objectives into the nosepiece in increasing order of magnification. Ensure they are securely tightened.

5. Connexió d'alimentació:

Connect the power cord to the microscope's power input and then to a suitable electrical outlet (90-240V AC).

lateral view of the AmScope T720 Trinocular Compound Microscope

lateral view of the AmScope T720 Trinocular Compound Microscope, highlighting the ergonomic design of the stand and the position of the coarse and fine focus knobs.

Instruccions de funcionament

1. Power On and Illumination:

Turn on the main power switch, usually located at the base of the microscope. Adjust the brightness control knob to achieve comfortable illumination. The Kohler illumination system provides even and bright lighting for optimal viewing.

2. Specimen Placement:

Col·loqueu el portaespècimens preparat sobre el suport mecànic.tage. Utilitzeu la stage clips to hold the slide securely in place. Use the X-Y stage control knobs to center the specimen area you wish to observe under the objective lens.

3. Enfocament:

  1. Enfocament aproximat: Begin with the 4X objective. Use the coarse focus knob to bring the specimen into approximate focus.
  2. Enfocament precís: Once roughly focused, use the fine focus knob for precise focusing. This knob allows for minute adjustments to achieve a sharp image.
  3. Interpupillary Distance and Diopter Adjustment: Adjust the interpupillary distance by moving the eyepiece tubes until a single, circular field of view is observed. If one eye requires a different focus, use the diopter adjustment ring on one of the eyepieces to compensate.

4. Changing Magnification:

To increase magnification, rotate the revolving nosepiece to the next higher objective (e.g., from 4X to 10X). Due to the Plan Infinity optical system, the image should remain largely in focus, requiring only minor adjustments with the fine focus knob. When using the 100X objective, apply a small drop of immersion oil directly onto the specimen slide before rotating the 100X objective into position. Ensure the objective lens makes contact with the oil.

5. Using the Trinocular Port:

The trinocular port allows for the attachment of a microscope camera or other imaging device. To direct light to the trinocular port, pull out the light path selector rod (if present) on the side of the head. Refer to your camera's manual for specific attachment and operation instructions.

Primer pla view of the AmScope T720 microscope's objectives and mechanical stage

Primer pla view of the AmScope T720 microscope's revolving nosepiece with Plan Achromatic objectives (4X, 10X, 40X, 100X) and the mechanical stage, illustrating the ample working distance provided by the backward nosepiece.

Manteniment

1. Neteja:

2. Emmagatzematge:

When not in use, cover the microscope with the provided dust cover to protect it from dust and debris. Store it in a cool, dry place.

3. Comprovacions periòdiques:

Periodically check all screws and moving parts for tightness. Do not overtighten.

Resolució de problemes

ProblemaCausa possibleSolució
Sense il·luminacióPower cord disconnected, power switch off, bulb failureCheck power connection, turn on switch, contact support for bulb replacement.
Image is blurry or won't focusIncorrect focusing, objective not fully engaged, dirty lens, slide upside downAdjust coarse and fine focus, rotate nosepiece until objective clicks into place, clean lenses, ensure slide cover slip is up.
Taques fosques al camp de viewDust on eyepiece, objective, or condenserClean eyepieces, objectives, and condenser. Rotate eyepiece to see if spots move.
Imatge massa fosca o massa brillantIllumination intensity incorrect, iris diaphragm setting incorrectAdjust brightness control, adjust iris diaphragm for optimal contrast.
Difficulty with 100X objectiveNo immersion oil, air bubbles in oil, incorrect oil typeEnsure a drop of immersion oil is used, remove air bubbles by gently rotating objective, use only specified immersion oil.

Especificacions

CaracterísticaDetall
ModelT720
Sistema òpticPlan Infinity
CapTrinocular, 30-degree inclined, 360-degree swiveling
OcularsWF10X (Extreme Widefield)
ObjectiusPlan Achromatic 4X, 10X, 40X (Spring), 100X (Spring, Oil)
Ampliació40X, 100X, 400X, 1000X
NasalBackward, Revolving
Stage3D Mechanical Stage
Centrant-seCoaxial Coarse & Fine Adjustment
Il·luminacióKohler LED, Wide Band Voltage (90-240 V)
Tipus de font de llumLED
MaterialGlass (Optical Components)
Pes de l'article28 lliures (12.7 kg)
Dimensions del paquet22.44 x 20.87 x 10.24 polzades

Garantia i Suport

For warranty information, technical support, or service inquiries, please contact AmScope customer service. Refer to the official AmScope website or your purchase documentation for the most current contact details and warranty terms.

També poden estar disponibles recursos en línia i preguntes freqüents a AmScope Store.

Documents relacionats - T720

Preview Manual d'usuari del microscopi estèreo AmScope sèrie PM240
Manual d'usuari per als microscopis estereoscòpics AmScope de la sèrie PM240, que cobreix la configuració, el funcionament, les especificacions i la resolució de problemes per a models binoculars i trinoculars.
Preview Manual d'usuari del microscopi AmScope sèrie 120
Manual d'usuari complet per als microscopis AmScope de la sèrie 120, que cobreix la configuració, el funcionament, les especificacions i la resolució de problemes per a models com el B120 i el T120.
Preview Manual d'usuari del microscopi AmScope sèrie M150
Manual d'usuari per als microscopis AmScope de la sèrie M150, que cobreix la configuració, el funcionament, el manteniment, les especificacions i la resolució de problemes.
Preview Manual d'usuari del microscopi compost digital Full HD AmScope DM150-W
Manual d'usuari complet per al microscopi compost digital Full HD AmScope DM150-W, que cobreix la configuració, el funcionament, la seguretat i el manteniment per a un ús òptim.
Preview Manual de l'operador del microscopi compost de la sèrie AmScope 490
Aquest manual d'instruccions proporciona instruccions completes per al microscopi compost AmScope sèrie 490, que cobreixen la configuració, el funcionament, el manteniment i la resolució de problemes per a un ús òptim.
Preview Manual de l'operador del microscopi compost de la sèrie AmScope 490 | Guia de l'usuari
Aquest manual d'instruccions proporciona instruccions completes per al microscopi compost AmScope sèrie 490, que cobreixen la configuració, el funcionament, el manteniment, la resolució de problemes i les especificacions. Apreneu a utilitzar el vostre microscopi AmScope de manera eficaç amb aquesta guia detallada.