Instruments Apera AI521

Apera Instruments AI521 PH800 Laboratory Benchtop pH Meter Kit Instruction Manual

Model: AI521 PH800

1. Producte acabatview

The Apera Instruments PH800 Benchtop pH Meter is engineered for precise and reliable pH measurements in various water solutions, suitable for scientific research and quality control applications. This kit includes Apera’s PC-body 201T-F pH/Temperature Electrode, enabling quick and accurate pH readings with automatic temperature compensation.

Característiques principals:

  • Quick and easy 1 to 3 points automatic calibration with a calibration guide, automatically recognizing 5 types of pH standards (7.00, 4.00, 10.01, 12.45, 1.68).
  • High-quality 201T-F pH/temperature combination electrode ensures stable and accurate measurements for general water solutions.
  • GLP data management with 500 groups of data storage and USB data output.
  • Electrode slope display between calibrations to monitor the health and performance of your pH electrode.
  • Large backlit LCD displays pH and temperature simultaneously, along with icons for stable readings and completed calibrations.
  • ORP (Redox) measurement mode available with an optional ORP electrode (sold separately).
  • IP54 splash-proof design for enhanced durability.

2. Contingut del paquet

The Apera Instruments PH800 Benchtop pH Meter Kit includes the following components:

  • pH-metre de sobretaula PH800
  • 201T-F pH/Temperature Electrode
  • Braç i suport d'elèctrode flexible
  • Cable USB
  • PC-Link Software Flash Drive (for Windows-based systems only)
  • DC9V (120V) Power Adapter
  • Calibration Solutions: 3 bottles (7.00, 4.00, 10.01, 50ml each)

3. Configuració inicial

Follow these steps to set up your PH800 Benchtop pH Meter for first use:

  1. Desempaquetar components: Traieu amb cura tots els articles de l'embalatge i comproveu si hi ha algun dany.
  2. Assemble Electrode Holder: Attach the flexible electrode arm and holder to the meter base. Ensure it is securely fastened to provide stable support for the electrode.
  3. Connect pH Electrode: Plug the BNC connector of the 201T-F pH/Temperature Electrode into the corresponding BNC port on the back of the meter. Connect the temperature probe into the TEMP port.
  4. Connectar l'alimentació: Plug the DC9V power adapter into the meter's power input port and then into a standard electrical outlet.
  5. Connexió USB opcional: If you plan to use the GLP data management features, connect the USB cable from the meter to your computer. Install the provided PC-Link software from the flash drive (Windows only).
Apera Instruments PH800 Benchtop pH Meter with electrode and beaker

Figura 3.1: The Apera Instruments PH800 Benchtop pH Meter fully assembled with the electrode in a beaker, demonstrating a typical setup for measurement.

posterior view of Apera PH800 pH meter showing connection ports

Figura 3.2: posterior view of the PH800 meter, highlighting the REF, pH/mV, TEMP, USB, and DC 9V connection ports for various probes and power.

Flexible electrode holder for Apera PH800 pH meter

Figura 3.3: Close-up of the flexible electrode holder, designed for easy positioning and secure support of the pH electrode during operation.

4. Instruccions de funcionament

4.1. Encès / apagat

Premeu el botó PODER button (usually marked with a circle and vertical line icon) to turn the meter on or off. The large backlit LCD will illuminate upon startup.

De dalt a baix view of Apera PH800 pH meter control panel

Figura 4.1: De dalt a baix view of the PH800 meter's control panel, showing the power button, mode/setup, enter, unit, calibration, and measurement buttons.

4.2. Calibració

Accurate calibration is crucial for reliable pH measurements. The PH800 supports 1 to 3 points automatic calibration.

  1. Preparació de solucions tampó: Use fresh Apera calibration solutions (pH 7.00, 4.00, 10.01 are commonly included). Ensure the electrode and beakers are clean.
  2. Iniciar el calibratge: Premeu el botó CAL button. The meter will enter calibration mode.
  3. Calibratge del primer punt: Rinse the electrode with distilled water and immerse it in the pH 7.00 buffer solution. Stir gently and wait for the reading to stabilize. The meter will automatically recognize the buffer and confirm calibration.
  4. Second and Third Point Calibration (Optional): For higher accuracy, repeat the process with pH 4.00 and pH 10.01 buffer solutions. The meter will guide you through the process.
  5. Calibration Reminder: The meter features a calibration reminder function to ensure regular calibration for optimal performance.
  6. Estat de l'elèctrode: The self-diagnosis function provides slope data between calibration points, allowing you to monitor the condition and performance of your pH electrode.

4.3. Mesurament

After successful calibration, you can proceed with measurements.

  1. Esbandir l'elèctrode: Rinse the electrode thoroughly with distilled water before and after each measurement.
  2. Elèctrode d'immersió: Submergiu l'elèctrode a la sample solution. Ensure the sensing part of the electrode is fully submerged.
  3. Espereu l'estabilització: Allow the reading on the large backlit LCD to stabilize. A stable reading icon (often a smiley face) will appear when the measurement is ready. The display shows both pH and temperature simultaneously.
  4. Lectura de registres: Note down the pH and temperature values.
Apera PH800 pH meter large backlit screen displaying pH and temperature

Figura 4.4: The large backlit LCD screen of the PH800 meter, clearly displaying pH and temperature readings, along with stabilization and calibration success icons.

4.4. Data Management (GLP & USB Output)

The PH800 meter offers advanced data management capabilities.

  • Emmagatzematge de dades: The meter can store up to 500 groups of measurement data, adhering to GLP (Good Laboratory Practice) standards.
  • USB Data Output: Connect the meter to a Windows-based computer via the USB cable to transfer stored data.
  • Programari de PC: Use the provided PC-Link software to view, analyze, and export data to spreadsheets for further analysis.
  • Auto-Timing Data Log: Utilize the auto-timing data log mode for professional measurements requiring timed data collection.
Apera PH800 pH meter connected to a laptop for USB data communication

Figura 4.5: The PH800 meter connected to a laptop via USB, illustrating its GLP data management capabilities with auto/manual data logging and 500 sets of data storage.

5. Manteniment

Proper maintenance ensures the longevity and accuracy of your PH800 pH meter and electrode.

  • Neteja d'elèctrodes: Regularly clean the pH electrode with distilled water. For stubborn residues, use a mild cleaning solution recommended for pH electrodes.
  • Emmagatzematge d'elèctrodes: When not in use, store the pH electrode in its protective cap filled with a suitable electrode storage solution (e.g., 3M KCl solution). Never store the electrode dry or in distilled water.
  • Neteja del comptador: Netegeu la superfície del mesurador amb un drap suauamp cloth. Avoid using abrasive cleaners or solvents. The meter is IP54 splash-proof, but avoid direct water exposure.
  • Solucions de calibratge: Keep calibration solutions sealed when not in use and replace them regularly to ensure accuracy.

6. Solució De Problemes

This section addresses common issues you might encounter with your PH800 pH meter.

  • Lectures inexactes: Ensure the electrode is properly calibrated. Check the electrode's slope data in the self-diagnosis function; a low slope indicates an aging or faulty electrode. Clean the electrode thoroughly.
  • Resposta lenta: This can be due to a dry or contaminated electrode. Rehydrate the electrode in storage solution or clean it.
  • Sense pantalla/alimentació: Check the power adapter connection and ensure the outlet is functional.
  • Errors de calibratge: Ensure you are using fresh, correct buffer solutions. Follow the calibration steps precisely. The self-diagnosis function can help identify specific calibration issues.
  • Problemes de transferència de dades: Verify the USB cable connection and ensure the PC-Link software is correctly installed on a compatible Windows system.

For persistent issues, refer to the detailed troubleshooting guide in the full user manual or contact Apera Instruments customer support.

7. Especificacions tècniques

EspecificacióValor
Número de modelAI521
Precisió del pH0.01 pH ± 1 digit
Rang de pH-2.00 a 19.99 pH
Resolució de pH0.1/0.01 pH
Compensació de temperatura0 - 100°C Automatic
Emmagatzematge de dades500 grups (compatible amb GLP)
SortidaUSB
Tipus de connectorBNC estàndard
Protecció d'entradaIP54 a prova d'esquitxades
Dimensions del producte9.5 x 8 x 4 polzades
Pes del producte5.5 lliures
CertificacionsCE, RoHS, ISO 9001:2015

8. Garantia i Suport

The Apera Instruments PH800 Benchtop pH Meter is manufactured under an ISO 9001:2015 certified quality system and is CE and RoHS certified, ensuring high standards of quality and safety.

For specific warranty information, technical assistance, or service inquiries, please contact Apera Instruments directly through their official website or authorized distributors. Please have your model number (AI521) and purchase date available when contacting support.

Documents relacionats - AI521

Preview Manual d'instruccions del mesurador de pH/conductivitat de taula APERA INSTRUMENTS PC950
Manual d'instruccions complet per al mesurador de pH/conductivitat de taula APERA INSTRUMENTS PC950, que cobreix les especificacions, el funcionament, el calibratge, el manteniment i la resolució de problemes per a ús al laboratori.
Preview Apera TN500 Portable Turbidity Meter: Instruction Manual and Specifications
This document provides comprehensive instructions and technical specifications for the Apera TN500 Portable Turbidity Meter. Learn about its features, calibration, measurement procedures, data management, and maintenance for accurate water quality analysis.
Preview Manual d'instruccions del mesurador de pH de taula Apera PH700
Manual d'instruccions complet per al pH-metre de taula Apera PH700, que cobreix les característiques, les especificacions, el funcionament, la calibració, el manteniment i la resolució de problemes.
Preview Manual d'instruccions del mesurador de pH de taula Apera PH700
Instruccions completes per al funcionament, calibratge i manteniment del pH-metre de taula Apera PH700. Cobreix les especificacions tècniques, la descripció de l'instrument, els procediments de mesura de pH i ORP, la configuració dels paràmetres, la cura dels elèctrodes i la resolució de problemes.
Preview Manual d'usuari de l'elèctrode de pH LabSen 751 - Apera Instruments
Manual d'usuari de l'elèctrode de pH de llança amb funda d'acer inoxidable LabSen 751 d'Apera Instruments, que detalla les característiques, les especificacions tècniques, l'ús, el manteniment i la garantia per a les proves de pH dels aliments.
Preview Apera Instruments PC60 & PH20 Pocket Tester User Manuals
Comprehensive user manuals for Apera Instruments PC60 Premium Series 5-in-1 Multi-Parameter Pocket Tester and PH20 Value Series pH Tester Kit. Includes setup, calibration, measurement, troubleshooting, and specifications.