1. Introducció
The Walfront HT-1202 Digital Water Quality Tester is a versatile instrument designed for accurate measurement of pH and mV values, along with temperature, in various aqueous solutions. This device is suitable for a wide range of applications including environmental monitoring, wastewater treatment, chemical laboratories, pharmaceutical research, fermentation processes, chemical engineering, and water supply analysis. Its high measurement accuracy and user-friendly design make it an essential tool for professionals and enthusiasts alike.

Image 1.1: Walfront HT-1202 Digital Water Quality Tester
2. Característiques del producte
- Ampli rang de mesura: Measures pH from 0 to 14 with a 0.01 resolution and mV from -415mV to 415mV.
- Alta precisió: Provides pH measurements with an accuracy of up to ±0.05pH and mV measurements with ±0.1% FS.
- Compensació de temperatura: Features automatic temperature compensation for accurate readings across a range of 0℃ to 60℃.
- Pantalla LCD gran: Equipped with a large LCD that simultaneously displays pH or mV and temperature. The backlight activates automatically for clear viewen diferents condicions d'il·luminació.
- Apagat automàtic: Includes an automatic power-off function after 4 minutes of inactivity to conserve battery life.
- Comprovació de calibratge: Incorporates a calibration check function to validate instrument performance and ensure reliability.

Image 2.1: Key components of the HT-1202 Digital Water Quality Tester
3. Contingut del paquet
En obrir el paquet, comproveu que hi hagi tots els elements següents:
- 1 x Walfront HT-1202 Digital Water Quality Tester
- 1 x Combined Electrode
- 1 x Package of Standard Buffer Reagents (for calibration)
- 1 x Electrode Soaking Liquid
- 1 x manual d'instruccions
- 1 x Black Carrying Case

Image 3.1: HT-1202 tester and its accessories

Image 3.2: HT-1202 kit in its protective carrying case
4. Configuració
4.1 Instal·lació de la bateria
- Localitzeu la tapa del compartiment de la bateria a la part posterior del dispositiu.
- Feu lliscar o desenganxeu la tapa per obrir el compartiment.
- Insert the required batteries (typically AA, refer to the compartment label for exact type and polarity) ensuring correct orientation.
- Tanqueu bé la tapa del compartiment de la bateria.

Image 4.1: Battery compartment of the HT-1202
4.2 Connexió d'elèctrodes
- Carefully remove the protective cap from the combined electrode.
- Connect the electrode cable to the host interface port located at the top of the HT-1202 tester. Ensure a firm and secure connection.
4.3 Preparació inicial
- Electrode Soaking: Before first use or after prolonged storage, soak the electrode in the provided electrode soaking liquid for at least 30 minutes to ensure optimal performance.
- Buffer Solution Preparation: When preparing buffer solutions for calibration using the provided reagents, use distilled water that has been boiled for 15-30 minutes and then cooled. This process removes dissolved carbon dioxide, which can affect pH accuracy. Avoid prolonged exposure to air during cooling to prevent re-absorption of carbon dioxide.
5. Instruccions de funcionament
5.1 Encès/Apagat
- Per encendre el dispositiu, premeu el botó Poder button. The LCD backlight will automatically activate.
- The device will automatically power off after 4 minutes of inactivity to save energy. To manually power off, press and hold the Poder botó.
5.2 Calibració
Regular calibration is crucial for maintaining the accuracy of your HT-1202 tester. Use the provided standard buffer reagents to prepare calibration solutions (e.g., pH 4.00, pH 6.86, pH 9.18).
- Enceneu el comptador.
- Rinse the electrode with distilled water and gently blot dry with a clean tissue.
- Immerse the electrode into the first standard buffer solution (e.g., pH 6.86).
- Premeu el botó Menú button to enter calibration mode. Follow the on-screen prompts to select the appropriate buffer value.
- Premeu Entra to confirm. The meter will automatically recognize the buffer and calibrate. Wait for the reading to stabilize and for the calibration indicator to confirm completion.
- Repeat the process for other buffer solutions (e.g., pH 4.00 and pH 9.18) as needed for multi-point calibration.
- After calibration, rinse the electrode thoroughly with distilled water.
5.3 Mesurament
- Ensure the meter is calibrated and the electrode is clean.
- Submergiu l'elèctrode a la sample solution to be tested. Ensure the electrode tip is fully submerged.
- Gently stir the solution to ensure homogeneity and allow the reading to stabilize.
- Read the pH or mV value and temperature displayed on the LCD screen.
- Després de la mesura, esbandiu l'elèctrode amb aigua destil·lada i torneu a col·locar la tapa protectora.

Image 5.1: HT-1202 in operation
6. Manteniment
6.1 Cura dels elèctrodes
- Always keep the electrode tip moist. Store it in the provided electrode soaking liquid or a suitable storage solution when not in use. Never store the electrode dry.
- Rinse the electrode thoroughly with distilled water before and after each measurement to prevent cross-contamination and residue buildup.
- If the electrode becomes dirty or readings become sluggish, clean it according to the electrode manufacturer's instructions (typically involving a mild cleaning solution).
6.2 Substitució de la bateria
When the low power indicator appears on the screen, replace the batteries promptly to ensure continuous and accurate operation. Refer to Section 4.1 for battery installation instructions.
6.3 Emmagatzematge
Store the HT-1202 tester and its accessories in the provided black carrying case in a cool, dry place, away from direct sunlight and extreme temperatures. Ensure the electrode is properly stored with its protective cap and soaking solution.
7. Solució De Problemes
- No Power / Display Off: Comproveu la instal·lació de les piles i assegureu-vos que no estiguin esgotades. Substituïu-les si cal.
- Lectures inexactes:
- Recalibrate the meter using fresh standard buffer solutions.
- Ensure the electrode is clean and properly soaked.
- Verify that the temperature compensation is functioning correctly.
- Check for any damage to the electrode.
- Temps de resposta lent: The electrode may be dry or dirty. Clean and re-soak the electrode.
- Indicador de baixa potència: Substituïu les piles immediatament.
8. Especificacions
| Paràmetre | Valor |
|---|---|
| Measurement Range (pH) | pH de 0.00 a 14.00 |
| Measurement Range (mV) | -415mV - 415mV |
| Resolution (pH) | 0.01 pH |
| Resolució (mV) | 1 mV |
| Accuracy (pH) | ±0.05 pH |
| Accuracy (mV) | ±0.1% FS |
| Interval de compensació de temperatura | 0 ℃ - 60 ℃ |
| Measured Solution Temperature | 5 ℃ - 60 ℃ |
| Pantalla LCD | Three and a half digits |
| Apagat automàtic | 4 minuts |
| Dimensions del producte | 6.5 x 2.95 x 1.34 polzades (165 x 75 x 32 mm) |
| Pes | Approx. 1.54 Pounds (692 g) |
| Fabricant | Walfront |

Image 8.1: HT-1202 product dimensions
9. Environmental Conditions for Use
To ensure optimal performance and longevity of the instrument, adhere to the following environmental conditions:
- Temperatura ambient: 5 ℃ - 40 ℃
- Humitat relativa: ≤85% HR (sense condensació)
- Vibració: Operate in an environment free from vibrations that could affect performance.
- Gasos corrosius: Ensure the surrounding air is free from corrosive gases.
- Interferència magnètica: Avoid other magnetic field interference outside the Earth's magnetic field.
10. Garantia i Suport
For information regarding warranty coverage, technical support, or service inquiries for your Walfront HT-1202 Digital Water Quality Tester, please refer to the contact information provided with your purchase documentation or visit the official Walfront weblloc web. Conserveu el rebut de compra com a comprovant de compra per a qualsevol reclamació de garantia.