1. Introducció
This instruction manual provides essential information for the safe and efficient operation, maintenance, and troubleshooting of your new Thermo Scientific™ Touch Screen Dry Bath Incubator. Please read this manual thoroughly before operating the instrument to ensure proper usage and to prevent damage or injury. Keep this manual for future reference.
2. Informació de seguretat
Always observe the following safety precautions to ensure safe operation and to avoid personal injury or damage to the instrument.
- Ensure the instrument is placed on a stable, level surface away from flammable materials.
- Do not operate the dry bath with damaged power cords or if it has been dropped or damaged in any way.
- Desconnecteu sempre l'alimentació abans de netejar o realitzar tasques de manteniment.
- Use appropriate personal protective equipment (PPE) when handling hot blocks or samples.
- The instrument is equipped with an over-temperature protection function. Do not bypass or tampamb aquesta característica de seguretat.
- This product is intended for professional use in laboratory environments.
3. Producte acabatview
The Thermo Scientific™ Touch Screen Dry Bath Incubator is a versatile laboratory instrument designed for precise temperature control of samples. It features an intuitive touch screen display for easy operation and accommodates a variety of blocks for diverse applications, including PCR sampel pretractament.
Característiques principals:
- Interfície de pantalla tàctil fàcil d'utilitzar.
- Compatible with various block types, including those for qPCR tubes.
- Suitable for PCR sampel pretractament.
- Intuitive touch screen display for straightforward operation.
- Equipped with a safe over-temperature protection function.

Figure 3.1: Image of the Thermo Scientific Touch Screen Dry Bath Incubator. Please note that the image provided is a placeholder indicating that the actual product image is currently being prepared.
4. Configuració
- Desembalatge: Carefully remove the dry bath incubator from its packaging. Inspect for any signs of damage during transit.
- Col·locació: Place the unit on a stable, level, and heat-resistant surface. Ensure adequate ventilation around the unit. Avoid placing it near heat sources or in direct sunlight.
- Block Installation: Insert the desired heating blocks into the wells. Ensure they are seated properly for optimal heat transfer. The unit supports 4 blocks.
- Connexió d'alimentació: Connect the power cord to the dry bath incubator and then to a grounded electrical outlet. Ensure the power supply matches the instrument's requirements (100-120V, 50/60Hz).
- Encesa inicial: Turn on the main power switch, usually located at the rear or side of the unit. The touch screen display should illuminate.
5. Funcionament
5.1. Funcionament bàsic
- Encès: Ensure the unit is connected to power and switched on.
- Configuració de la temperatura:
- On the touch screen, navigate to the temperature setting menu.
- Use the up/down arrows or numeric keypad to set the desired temperature within the range of ambient +5°C to 130°C.
- Confirm the setting. The unit will begin heating to the set temperature.
- Configuració del temporitzador:
- Access the timer setting menu on the touch screen.
- Set the desired time from 0 to 99 minutes and 59 seconds, or select continuous operation.
- The timer will start once the set temperature is reached, or immediately if continuous operation is selected.
- Col·locació de Samples: Once the desired temperature is reached, carefully place your samples in the heating blocks.
- Seguiment: The touch screen will display the current temperature and remaining timer (if set).
- Finalització de l'operació: At the end of the timed cycle, the unit may beep or display a completion message. Remove samples carefully using appropriate tools. Turn off the unit if not needed for further use.
5.2. PCR Sample Pre-treatment
The dry bath incubator is ideal for various PCR sample pre-treatment protocols. Refer to your specific PCR protocol for recommended temperatures and incubation times. The precise temperature control and uniformity of this unit ensure reliable results for sensitive applications.
6. Manteniment
Regular maintenance ensures the longevity and optimal performance of your dry bath incubator.
- Neteja:
- Desconnecteu sempre l'alimentació abans de netejar.
- Allow the unit and blocks to cool completely.
- Netegeu les superfícies exteriors amb un damp drap i detergent suau. No utilitzeu netejadors abrasius ni dissolvents.
- Clean the wells and blocks as needed. Blocks can be removed for thorough cleaning.
- Substitució de fusibles: If the unit does not power on, check the fuse protector (250V 8A). Replace only with a fuse of the same type and rating. Consult a qualified technician if unsure.
- Emmagatzematge: When not in use for extended periods, store the unit in a clean, dry environment.
7. Solució De Problemes
Aquesta secció tracta els problemes habituals que podeu trobar. Per a problemes que no figuren aquí, poseu-vos en contacte amb l'atenció al client.
| Problema | Causa possible | Solució |
|---|---|---|
| La unitat no s'encén. | Sense subministrament d'alimentació; fusible fundit; interruptor apagat. | Check power cord connection; Replace fuse (250V 8A); Ensure power switch is ON. |
| Temperature not reaching set point. | Incorrect temperature setting; Poor block contact; Ambient temperature too low. | Verify set temperature; Ensure blocks are properly seated; Operate in a suitable ambient temperature. |
| La temperatura fluctua significativament. | Sensor issue; Unit malfunction. | Contact customer support for diagnosis and repair. |
| La pantalla tàctil no respon. | Temporary software glitch; Hardware issue. | Reinicieu la unitat; si el problema persisteix, poseu-vos en contacte amb el servei d'atenció al client. |
8. Especificacions
| Paràmetre | Valor |
|---|---|
| Número de model | 88870009 |
| Nombre de blocs | 4 |
| Interval de temperatura | Ambient +5°C to 130°C (at 25°C ambient) |
| Temperature Control Accuracy (at 37°C) | ±0.5 °C |
| Temperature Uniformity (at 37°C) | ±1 °C |
| Mètode de control de temperatura | PID Digital Control |
| Interval de configuració del temporitzador | 0 to 99 minutes 59 seconds or Continuous Operation |
| Temps d'escalfament (de 30 °C a 130 °C) | Menys de 25 minuts |
| Interval d'humitat relativa | Per sota del 80% |
| Font d'alimentació | 100-120V, 50/60Hz |
| Fuse Protector | 250V 8A |
| Dimensions (W×D×H) | 200×535×100 mm |
| Pes | 5.1 kg |
9. Garantia i Suport
Thermo Fisher Scientific warrants this product to be free from defects in material and workmanship under normal use and service. For specific warranty terms and conditions, please refer to the warranty card included with your product or visit the official Thermo Fisher Scientific website. For technical support, service, or to order replacement parts, please contact your local Thermo Fisher Scientific representative or authorized distributor. When contacting support, please have your model number (88870009) and serial number ready.
Informació de contacte:
- Weblloc: www.thermofisher.com
- Atenció al client: Refer to your regional Thermo Fisher Scientific contact details.