1. Introducció
This user manual provides detailed instructions for the operation, maintenance, and troubleshooting of the Irfora Bench Digital LCR Meter. This high-precision instrument is designed for accurate measurement of inductance (L), capacitance (C), and resistance (R) across a wide frequency range.
Featuring a 3.5-inch LCD screen and a user-friendly interface, the meter supports multiple test frequencies and levels, offering a basic accuracy of 0.1%. It includes USB and RS232 communication interfaces for remote operation and a handler interface for automated component sorting.
2. Informació de seguretat
WARNING: To prevent electric shock or damage to the instrument, please read and follow all safety instructions carefully before operation.
- Assegureu-vos la potència correcta voltage (AC 200-240V, 50/60Hz) is used as specified on the device.
- Do not operate the meter in wet or damp condicions.
- No obriu l'instrument casing; no hi ha peces a l'interior que l'usuari pugui reparar. Demaneu totes les tasques de manteniment a personal qualificat.
- Always disconnect the power cord before cleaning or moving the instrument.
- Use only the provided accessories (Kelvin Alligator Clip, power cord, fuses).
- Avoid exposing the instrument to extreme temperatures or direct sunlight.
3. Producte acabatview
The Irfora Bench Digital LCR Meter is designed for precision and ease of use. Below are the key components and features of the device.

Figura 3.1: Frontal view of the Irfora Bench Digital LCR Meter, showing the display and control panel.
3.1 Disposició del panell frontal

Figura 3.2: Detallada view of the front panel with numbered components for identification.
| Número | Descripció |
|---|---|
| 1 | L’interruptor d’alimentació |
| 2 | Pantalla LCD de 3.5 polzades |
| 3 | Grounding and test terminals (H Force, H Sense, L Sense, L Force) |
| 4 | Basic function keys (LEVEL, FREQ, RANGE, SPEED, ANULL, CAL, SET, ENTER, AUTO/ICLZ, COMPARE, SAVE, TOL%, HOLD, DCR, MAX, MIN, AVG) |
3.2 Disseny del panell posterior

Figura 3.3: Detallada view of the rear panel with numbered components for identification.
| Número | Descripció |
|---|---|
| 1 | Power supply plug base and fuse case |
| 2 | Voltage selector |
| 3 | Interfície RS232 |
| 4 | Interfície de dispositiu USB |
| 5 | Interfície de manipulador |
3.3 Característiques de visualització

Figure 3.4: The 3.5-inch LCD screen provides clear, 5-digit measurement readings and parameter settings.
The 3.5-inch LCD screen displays measurement results, selected parameters (e.g., L, C, R, Z, X, D, Q, θ, ESR), test frequency, test level, bias voltage, and other operational settings. The 5-digit display ensures high resolution for precise readings.
3.4 Mànec ajustable

Figure 3.5: The integrated handle can be adjusted to various positions, allowing for flexible viewing angles and improved usability on a workbench.
4. Configuració
Follow these steps to set up your Irfora Bench Digital LCR Meter for first use.
- Desembalatge: Carefully remove the LCR meter and all accessories from the packaging. Verify that all items listed in the package list are present: Bench Digital LCR Meter, Kelvin Alligator Clip, 2 Fuses, Power Cord, User Manual (English).
- Connexió d'alimentació:
- Assegureu-vos que el voltage selector (2 on Figure 3.3) on the rear panel is set to the correct voltage for your region (AC 200-240V).
- Connect the provided power cord to the power supply plug base (1 on Figure 3.3) on the rear panel of the meter.
- Connecteu l'altre extrem del cable d'alimentació a una presa de corrent amb connexió a terra.
- Encesa inicial: Press the power switch (1 on Figure 3.2) located on the front panel to turn on the device. The LCD screen (2 on Figure 3.2) will illuminate and display the startup sequence.
- Connexió dels cables de prova: Connect the Kelvin Alligator Clip to the grounding and test terminals (3 on Figure 3.2) on the front panel. Ensure a secure connection for accurate measurements.
5. Instruccions de funcionament
This section details the basic operation and advanced functions of the LCR meter.
5.1 Mesura bàsica
- Encès: Turn on the LCR meter using the power switch.
- Component de connexió: Connect the component to be measured to the Kelvin Alligator Clip.
- Select Measurement Function: Use the function keys (4 on Figure 3.2) to select the desired measurement parameter (L/C/R/Z). The primary and secondary parameters (X/D/Q/θ/ESR) will be displayed automatically.
- Ajust de freqüència: Premeu el botó FREQ button to cycle through available test frequencies (100Hz, 120Hz, 200Hz, 400Hz, 800Hz, 1KHz, 2KHz, 4KHz, 8KHz, 10KHz, 15KHz, 20KHz, 40KHz, 50KHz, 80KHz, 100KHz).
- Adjust Test Level: Premeu el botó NIVELL button to select the desired test signal level (100mV, 300mV, 600mV, 1000mV, 1500mV, 2000mV).
- Llegir mesura: The measurement results will be displayed on the 3.5-inch LCD screen.
5.2 Funcions de correcció
The meter supports Open Circuit Correction and Short Circuit Correction to improve measurement accuracy, especially for low-value components or long test leads.
- Open Circuit Correction: Sense cap component connectat als terminals de prova, premeu el botó CAL button and follow the on-screen prompts for open circuit correction.
- Short Circuit Correction: Short the test terminals together (e.g., using a shorting bar or by touching the alligator clips together), then press the CAL button and follow the on-screen prompts for short circuit correction.
5.3 Screening and Comparator Functions
The screening function allows setting tolerance levels for component sorting, while the comparator function enables automatic sorting into qualified and failed bins.
- Tolerance Setting: Utilitza el TOL% button to set tolerance levels from -100% to 100%, with fixed points at 10%, 25%, and 50%.
- Comparator Sorting: The comparator function offers 5 gear sorting, including 3 qualified gears, 1 failed gear, and 1 ancillary gear. This is particularly useful when using the Handler interface for automated testing.
5.4 Interfícies de comunicació
The meter is equipped with USB, RS232, and Handler interfaces for remote control and data acquisition.
- Interfície USB: Connect the meter to a computer via the USB port (4 on Figure 3.3) for data logging and remote control using compatible software.
- Interfície RS232: The RS232 port (3 on Figure 3.3) allows for serial communication with external systems.
- Handler Interface: The Handler interface (5 on Figure 3.3) is designed for integration into automated test systems, facilitating rapid component sorting based on set criteria.
6. Manteniment
Proper maintenance ensures the longevity and accuracy of your LCR meter.
- Neteja: Use a soft, dry cloth to clean the exterior of the instrument. For stubborn dirt, a slightly damp Es pot fer servir un drap amb detergent suau, però assegureu-vos que no entri líquid al dispositiu. No utilitzeu productes de neteja abrasius ni dissolvents.
- Emmagatzematge: When not in use, store the meter in a clean, dry environment away from direct sunlight, dust, and extreme temperatures.
- Substitució de fusibles: If the meter does not power on, check the fuse located near the power input on the rear panel (1 on Figure 3.3). Replace with a fuse of the same type and rating (AC 250V 500mA). Always disconnect the power cord before replacing the fuse.
- Calibració: Regular calibration by qualified personnel is recommended to maintain the instrument's specified accuracy.
7. Solució De Problemes
Aquesta secció ofereix solucions a problemes comuns que podeu trobar.
| Problema | Causa possible | Solució |
|---|---|---|
| El comptador no s'encén | No power supply; Blown fuse; Power cord not connected properly | Check power outlet; Replace fuse (AC 250V 500mA); Ensure power cord is securely connected |
| Lectures inexactes | Test leads not properly connected; No Open/Short correction performed; Component damaged; Incorrect test frequency/level selected | Ensure secure connection of Kelvin clips; Perform Open/Short correction; Test with known good component; Adjust frequency/level as per component type |
| La pantalla mostra un missatge d'error | Measurement range exceeded; Internal error | Adjust measurement range; Refer to specific error code in full manual (if applicable); Restart the device |
| Communication issues (USB/RS232) | Incorrect cable; Driver not installed; Software settings incorrect | Use correct cable; Install necessary drivers; Verify software communication settings |
Si el problema persisteix després d'intentar aquestes solucions, poseu-vos en contacte amb l'atenció al client.
8. Especificacions
Detailed technical specifications for the Irfora Bench Digital LCR Meter.

Figure 8.1: Dimensions of the LCR meter: 305mm (12.01in) length, 265mm (10.43in) width, 105mm (4.13in) height.
| Paràmetre | Valor |
|---|---|
| Número de model | LAKUSE20198EU |
| Pantalla | Pantalla LCD de 3.5 polzades |
| Dígits | 5 dígits |
| Paràmetres principals | L, C, R, Z |
| Paràmetres secundaris | X, D, Q, θ, ESR |
| Inductance (L) Range | 0.001µH ~ 9999H |
| Capacitance (C) Range | 0.001 pF ~ 9999 µF |
| Resistance (R) Range | 0.0001Ω ~ 99.99MΩ |
| Precisió bàsica | 0.1% |
| Excitation Source Frequency | 16 Points (100, 120, 200, 400, 800, 1K, 2K, 4K, 8K, 10K, 15K, 20K, 40K, 50K, 80K, 100K Hz) |
| Velocitat de visualització de mesura | 2 times/s (slow), 5 times/s (medium), 10 times/s (fast) |
| Biaix intern | 0-1500mV Adjustable, 1mV Stepping |
| Test Level (Vrms) | 100mV, 300mV, 600mV, 1000mV, 1500mV, 2000mV |
| Funció de correcció | Open Circuit Correction, Short Circuit Correction |
| Screening Function | Tolerance -100%~100%, with fixed points of 10%, 25%, 50% |
| Comparator Selection | 5 Gear Sorting (3 Qualified, 1 Failed, 1 Ancillary) |
| Interfície de comunicació | USB, RS232, Handler |
| Temperatura de treball | 0 ~ 40 ℃ |
| Humitat de treball | 15-85% HR |
| Mida de l'article | 305 x 265 x 105mm (12.01 x 10.43 x 4.13 polzades) |
| Pes de l'article | 2111 g (4.65 lliures) |
| Voltage | CA 200-240V, 50/60Hz |